11. BUILT-IN DISPLAY FUNCTIONS AND OPERATION
(Totalizer or indicator equipped model)
Description of Display Functions
This totalizer can display a total of four different readings: total flow, actual instantaneous flowrate,
percent instantaneous flowrate, and 8-division % bar graph. It also shows the following error messages:
Full scale is exceeded: ErrorFS
Upper-end flowrate is exceeded: ErrorOF
NOTE:
When both errors above have occurred, message "ErrorOF" has priority over the other.
Display is selectable either with the display select
switch located inside the preamplifier or through
communications using the Smart Communication
Unit.
If communications is your choice, follow the
instructions outlined in the Smart Communication
Unit EL230 instruction manual.
NOTE: Select your option at "Preamplifier
Information" menu at "Setup" on the
PC screen.
With display select switch
Opening up the cover facing the internal assembly
of the preamplifier, press the display select switch
SW located on the isolation board. The display
will then scroll forward each time this switch is
pressed as shown in Fig. .2.
11.2 Total Flow Reset
Displayed total flow can be reset either by the display select switch SW or through communications
with Smart Communication Unit. If communication is your option, see the instruction manual for Smart
Communication Unit EL230.
NOTE : Select your option at "Measurement" at "Display" menu on the PC screen.
With Display Select Switch (totalizer equipped model)
While the totalizer is in the total flow display mode, holding the display select switch SW
depressed for more than 3 seconds resets the total flow.
11.1 Display Selection
(totalizer equipped model)