11.8 Error code
If an alarm mark (Fig.11-6) is displayed on the display of display unit, operate
switch a few times (max.
twice) to shift to the error code display screen. Respond to the error according to the code referencing the table
shown below.
SELECT
Fig. 11.6 Alarm mark
Code
Description
Response
0002
EEPROM data reading / writing error
- Turn OFF the power supply and restart the
instrument
- Initialize the parameter *
0004
Initialization error
Turn OFF the power supply once and restart
the instrument
0016
Failure of board (BASE) in display unit
Replace the set of display unit as a new one
0032
Failure of board (INTERFACE) in display unit
Replace the set of display unit as a new one
0128
Meter coefficient input error
Change the meter coefficient of parameter
setting to an appropriate value (for the
method for changing it, refer to section 10)
NOTE: * If the initialization is implemented, the measurement value may change. For the method for
initializing the parameter, please contact our company's sales department.
If the error cannot be reset by response shown above (excluding 0016, 0032), the display unit may be broken.
Please replace the set of display unit as a new one. (To purchase the replacement item, please contact OVAL's
sales department.)
12. Explosionproof structure
Temperature of fluid to be measured Ambient temperature
Class
Name of model
Depending on flowmeter
-20°C to +55°C
II CT6
EL0137-1