E-126-3-E
12
Union coupling
G3/4
Rubber packing
Screw M4
M25
Clamp
Gland
Fig. 5.1 External wiring gland assembly diagram
◆
Precautions on external wiring gland (pressure tight packing)
◆
For both of non-explosionproof type and explosionproof type, take the following into consideration.
1. The pressure-tight packing must fit the finished cable external diameter.
2. Securely tighten the pressure-tight packing so that there is no gap between the packing and the
cable.
3. If not using a pressure-tight packing, make sure to take necessary measures to prevent moisture and
dust from entering.
CAUTION: 1. After the cable is passed through the rubber packing, crush the screwed
packing of gland. At this time, screw the gland until the cable stop slipping.
2. This cable gland is a cable grand with anchor function. After the step(1), be
sure to screw the clamp with screw: M4 and fix the cable surely.
5.1 Close-up plug
When the close-up plug is used for TIIS explosionproof type, be sure to use the close-up plug designated by us.
Under the environment where dust and water immersion are concerned, use the plug designated by us to satisfy
the IP class of the product. (This instrument is shipped with the close-up plug sufficiently secured to ensure the
airtightness.)
5.2 Replacement of external wiring gland and close-up plug
External wiring gland and close-up plug are designated by us. If they need to be repaired or replaced, the
replacement parts can be procured from OVAL's agent in your area or OVAL service department. When you
order the parts, please inform us of flowmeter model, part name, and quantity.