E-025-2-E
2
④ The message box (Fig. 47) is displayed after changing the setting values. Click "OK" to complete
the setting process.
Fig.47
⑤ The window for setting item is displayed. To exit, click "Cancel".
3.6.3.5 Output at Error/Alarm (Error output)
① Click "Setup (S)" from the menu, and select "Error output" from the drop-down list of "Outputs",
then the following window (Fig. 48) will be displayed.
Fig.48
② Select for respective items.
Setting item
Detail
Setting range
Analog error
output
Analog output at error (Select from the right column.)
No Function: Normal output
Downscale: Hold by outputting 2.4mA
Zero: Hold by outputting 4.0mA
Hold: Hold current value just before occurrence of error
Upscale: Hold by outputting 21.6mA
No Function, Downscale,
Zero, Hold, Upscale
Pulse error
output
Pulse output at error (Select from the right column.)
No Function: Normal output
Zero: 0Hz (Stops pulse output)
Hold: Hold the frequency just before occurrence of error
Upscale: Outputs the frequency equivalent to 110% of
metering range.
No Function, Zero, Hold,
Upscale
NOTE: "Error output" functions when any one of "Overflow Error", "Hardware Failure",
"Parameter Alarm", "Self Diag Alarm", and "Empty Pipe Alarm" occurs. For errors,
refer to "3.11.4 Error/Alarm, Status List".