E-025-2-E
25
3.6.3.3 Status Output (Status output)
① Click "Setup (S)" from the menu, and select "Status output" from the drop-down list of "Outputs",
then the following window (Fig. 42) will be displayed.
Fig.42
③ Click "OK" after the setting is completed to display the message box (Fig. 43). Click "OK" here
to change the settings to the inputted values, and to reflect the changed setting values in the
flowmeter's output. If the flowmeter's output is used to control valves or other such parts as safety
protocol, change the control loop to manual control so that the control loop cannot be influenced
by the flowmeter's output.
② Select for respective items.
Setting item
Detail
Setting range
Status output func
Status output selection (Select from the right column.)
No Function: No function, Contact output OFF
Error Status: ACTIVE when relevant error occurs with
Status select
H/L Alarm: ACTIVE at high/low alarm
Flow Direction: ACTIVE at the occurrence of reverse flow
No Function, Error
Status, H/L Alarm,
Flow Direction
Status output mode
Status output mode (Select from the right column.)
Off Active: Output transistor Off at operation
On Active: Output transistor On at operation
Off Active, On Active
Fig.43
④ The message box (Fig. 44) is displayed after changing the setting values. Click "OK" to complete
the setting process.
Fig.44
⑤ The window for setting item is displayed. To exit, click "Cancel".