L-740FF-1-E
5. Explanation of Blocks
This product has following blocks:
・
Resource block
・
Transducer block
・
AI (analog input) function blocks: 4 (measurement of volume flow, mass flow, density, and temperature)
・
IT (integration) function blocks: 2 (volume flow and mass flow)
Following contents describe functions and handling of each block installed on the product.
5.1 Resource Block
This block contains information specific to fieldbus device such as device name, manufacturer name, and
serial number.
4. Switches on Communication Module (SIM_ENABLE Hard Switch)
DIP switches SW2 on communication module (FB-B board) are assigned with following functions:
Switch No.
Function
Description
1
unused
——
2
SIM_ENABLE
Set to "ON" when activating SIMULATE function of AI function
block or FD_SIMULATE function of resource block.
3
unused
——
4
unused
——
Functions are activated by placing each switch in "ON" position. It is recommended that
switches be placed in "OFF" position unless it is necessary.
Location of DIP switch SW2
Transmitter
Open this cover and remove the black resin cover
inside to expose a printed-circuit board for communi-
cation module.
SW2 is located on that printed-circuit board.
DIP SW2
Fig.4.1
NOTE: