L-737-3-E
9.11 View Variables Screen Setup
Display / Hide and priority of variables view can be set up at the view variables screen.
Follow the procedure outlined below:
() Bring up the item of function you want to use.
Firstly, proceed key operation to show the variables view priority setup (Var. Priority)
screen.
(For the procedure to view individual items, see .. thru ..7 Transition chart.)
When the cursor is placed over variables view priority setup (Var. Priority), click "SEL"
to bring up Var. Priority setup screen.
Click "SEL".
() Decide whether to change priority of variables or not.
The current priority of variables is shown.
If you want to change the priority, click "SEL3" and the next screen, Var. Priority,
appears.
If you want to cancel the priority change process, click "ENT" followed by clicking
"SEL3" and the window returns to menu select screen.
(3) View priority setup
The current view variables priority is shown.
Individual items are numbered on the right side according to their view priority. If
0 is set, the affected item disappears from the list. By entering numbers desired
and then scrolling through the list downward, the order of items are rearranged.
(For the setup procedure, see .3 Parameter Value Setup (3).)
Available items subject to priority rearrangement are
. Mass Flow: Instant mass flowrate
. Vol Fiow:
Instant volume rate or instant volume rate (fixed density)
3. Density:
Density
. Temp:
Temperature
. Counter:
Counter reading of pulse output
. Counter:
Counter reading of pulse output
7. Tortalizer: Total flow of pulse output
. Totalizer:
Total flow of pulse output
. Ana Out:
% reading of analog output
0. Ana Out:
% reading of analog output
Upon completion of setup, move the cursor to "Enter" and click "S"
A Var. Priority confirmation screen appears.
() Finalizing priority
①
If the priority is OK, click "SEL3".
Setup is complete and the window returns
to the initial menu select screen.
If you want to abort the setup, click "ENT"
and the screen below appears.
②
If you want to abort setup, click "SEL3".
Setup is complete and the window returns
to the initial menu select screen.
Click "SEL3".
Click "ENT".
Click "ENT".
Click "SEL3".
Click "SEL".
Click "ENT".
Click "ENT".
Click "SEL3".
Click "SEL3".
NOTE: For this function, the setting change is disabled
in Write Protect mode (refer to section .).