![Oval ALTImass CA004 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/oval/altimass-ca004/altimass-ca004_instruction-manual_1662154021.webp)
L-737-3-E
A
B
C
A.B
B.C
A
C
⑥
For the display unit that has just been removed, it is necessary to take off fitting screws once, adjust the
orientation and re-install them. Shown below are the location of screw holes for different orientations.
A
Liquid service,
horizontal run
B
Vertical run
C
Gas service,
horizontal run
Set washers (3 places)
LCD fitting screws
(M., 3 places)
Fig..
⑦
Align connector pins (male) on the transmitter with
mating through holes (female) in the display unit.
⑧
Tighten screws (3 places) at new screw holes and re-
install into the original position.
Connector
pins
Display
unit
through
holes
CAUTION: When tightening the display lid, tighten securely using a flat tool or similar
instrument. Insufficient tightening can affect gas-tightness and sensitivity of
display,s optical switch.
Fig..
Fig..7
Depending on
the transmitter
type, shapes of
the connector or
peripheral parts
may be different.