66 | Ottobock
Beep signal
Vibration
signal
Additional action per
formed
Event
3x short
3x short
Prosthetic leg moved next to
contralateral leg, set down
and kept still for 1 second
Switching to MyMode 2 (mode 3)
carried out.
13.2.2 Warnings/error signals
Error during use
Beep signal
Vibration signal
Event
Required action
1 x long at interval of
approx. 5 seconds
Overheated hydraulic
unit
Reduce activity.
3x long
Battery charge level
under 25%
Charge battery soon.
5x long
Battery charge level
under 15%
Charge battery promptly,
since the product is
switched off after the next
warning signal.
10x long
10x long
Battery charge level
0%
After the beep and
vibration signal, the
product switches to
empty battery mode
and then switches
off.
Charge the battery.
30x long
1x long, 1x short
repeated every 2.6
seconds
Severe error
e.g. failure of valve
drives
Possibly no switching
into safety mode.
Restricted walking pos
sible.
Attempt to reset this error
by connecting/discon
necting the battery char
ger. The battery charger
must be connected for at
least 5 seconds before it
is disconnected.
If the error persists, it is
no longer permissible to
use the product. The
product must be inspec
ted immediately by a
prosthetist.
Error while charging the product
LED on
power
supply
LED on bat
tery charger
Error
Resolution
Country-specific plug adapter not
fully engaged on power supply
Check whether the country-spe
cific plug adapter is fully
engaged on the power supply.