31
CAUTION
Unintentional unlocking of the Axon-Bus gripping component
Injury due to releasing the Axon-Bus gripping component from the forearm (e.g. while carrying
objects).
►
The two release buttons should only be activated in order to change the Axon-Bus gripping
component, deliberately and with due consideration of the respective situation.
CAUTION
Risk of pinching between the hook tips
Injury due to pinching of body parts.
►
Ensure that no body parts are between the hook tips when using or rotating the product.
CAUTION
Gripping objects with different gripping force when using the "AxonHook" gripping
component
Injury due to unexpected behaviour of the gripping component.
►
Note that the gripping force can vary depending on the consistency (soft/hard) of the object
being gripped.
CAUTION
Use of the Axon-Bus prosthetic system with dirty or damaged contacts of the rotation
unit and gripping component
Injury due to unexpected behaviour of the Axon-Bus prosthetic system.
►
Make sure that the contacts of the rotation unit and gripping component are always clean
and free of grease.
►
Clean the electrical contacts regularly using cotton swabs and a mild soap solution.
►
Take care to avoid damaging the contacts with pointed or sharp objects.
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
►
Only clean the product with a damp cloth and mild soap (e.g. 453H10=1-N Ottobock Der
maClean).
►
Use only the following products for cleaning/disinfecting the inner socket:
Cleaning:
453H10=1 Ottobock DermaClean
Disinfection:
Colourless, conventional medical disinfectants
4.6 Information on the Power Supply/Battery Charging
CAUTION
Charging the Axon-Bus prosthetic system with soiled or damaged contacts
Injury due to unexpected behaviour of the Axon-Bus prosthetic system caused by insufficient
charging.
►
Make sure that the contacts are always clean and free of grease.
►
Clean the electrical contacts of the charging plug and charging receptacle regularly using
cotton swabs and a mild soap solution.
►
Take care to avoid damaging the contacts with pointed or sharp objects.
Содержание Axon-Bus
Страница 2: ...1 2 3 4 5 6 2...
Страница 261: ...261 2 2 3 Axon Bus 757B500 757B501AxonEnergy Integral LED Axon Bus 2 2 4 9S501 360 24 15 9S503 160 2 2 5 8E500...
Страница 262: ...262 3 2 mm...
Страница 263: ...263 3 3 3 mm 0 12 8E600 2 1 1 2 3 3 3 1 3 2 1 274 3 3...
Страница 264: ...264 3 4 3 5 Ottobock 4 4 1 4 2 1 1 2 4 3...
Страница 265: ...265 AC AC 4 4...
Страница 266: ...266 CT MRI WiFi 30 cm 274 4 5...
Страница 267: ...267 Axon Bus Axon Bus Axon Bus 453H10 1 Ottobock 453H10 1 Ottobock 453H16 Ottobock...
Страница 268: ...268 4 6 4 7 12K501 12K500...
Страница 269: ...269 5 5 1 1 1 757L500 1 5 2 4X500 757S500 6 1 3 757L500 6 1 1 1 2 2 3 AC 6 2 1 278 2 3 6 3 LED 3 2 5 1 1 2 LED...
Страница 270: ...270 6 4 7 7 1 1 1 7 2 1 3 2 7 3 1 2...
Страница 271: ...271 1 1 2 4 3 7 4 1 5 6 2 1 5 6 2 9S501 9S503...
Страница 272: ...272...
Страница 277: ...277 PPPP YYYY WW PPPP YYYY WW 12 2 12 2 1 1 2 1 2 1 12 2 2 1 3...
Страница 278: ...278 12 2 3 AC 25 50 75 100 100 50...
Страница 282: ...282...
Страница 283: ......