Ottobock | 19
4R120, 4R121=30, 4R121=34
1) Grease the selected Elastomer Plate.
2) Arch the Elastomer Plate slightly (Fig. 8) and place it between the rotational section (Fig. 7, Pos.
5) and the stop (Fig. 7, Pos. 6).
3) Press the Elastomer Plate against the inner surface of the housing until the Elastomer Plate lines
up with the rotational section (Fig. 9).
NOTE: Insertion of the Rotational Locking Plate, durometer EXTRA-HARD, is made easier during
assembly if the black Elastomer Insert is inserted last between the white rotational section and the
black locking section. Disassembly is made easier if the black Elastomer Insert is removed first.
4) Press the inner part of the housing up to the stop
5) Set the DeltaTwist on a clean surface and hold the housing with one hand. While that hand presses
the housing downwards so that the blue retaining screw can be reached, turn the blue retaining
screw clockwise with the other hand (Fig. 10).
6) Tighten the blue retaining screw hand tight (approx. 4 Nm).
7) Insert the Threaded Bushing with Inner Friction Bearing and tighten it to between 15-20 Nm with
the 709H4 Hook Wrench. To apply the correct torque use the Pyramid Receiver, as shown in Fig.
11, where a prosthetic foot is used as an example.
8) Reset the tension of the Elastomer Rod by turning the red tension screw (clockwise).Make sure not
to loosen the blue locking screw while doing this.
9 Technical data
Article number
4R120
4R121=30
4R121=34
Proximal connection
Tube clamp
⌀ 30 mm
Tube
⌀ 30 mm
Tube
⌀ 34 mm
Distal connection
Pyramid receiver
Pyramid receiver
Pyramid receiver
Outer diameter
47 mm
47 mm
47 mm
Weight
~ 340 g
~ 530 g *
~ 585 g *
System height
117 mm
–
–
Min. system height
–
177 mm
218 mm
Max. system height
–
553 mm
578 mm
Material
Aluminum
Aluminum
Aluminum
Max. torsion inner/ outer
± 20°
± 20°
± 20°
Max. dampening
~ 8 mm
~ 8 mm
~ 8 mm
Mobility Grade
2 - 4
Max. body weight
100 kg
125 kg
10 Legal information
10.1 Liability
The manufacturer recommends using the device only under the specified conditions and for the
intended purposes. The device must be maintained according to the Instructions for Use. The
device must only be operated with tested modular components in accordance with the Ottobock
Mobility System. The manufacturer is not liable for damage caused by component combinations
that were not authorised by the manufacturer.