Parameter
Adjustment soft
ware range
Setting range,
app
Meaning
Pitch
1000 Hz
–
4000 Hz
1000 Hz
–
4000 Hz
Pitch of beep signal for confirmation tones
Volume
0 – 4
0 – 4
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch
ing MyModes). The "0" setting deactivates
the audible feedback signals. However,
warning signals in case of errors are still gen
erated.
9.2.3 Overview of adjustment parameters in MyModes
CAUTION
Improper use of the setting parameters in the MyModes
Falling due to unexpected product behaviour because of changed damping behaviour.
►
Instruct the patient regarding the functionality and adjustment options for
all parameters
of the MyModes.
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
The parameters in the MyModes describe the static behaviour of the prosthesis for a specific movement pattern
such as cycling. Damping behaviour is not automatically controlled and adjusted in MyModes.
Parameter
Adjustment soft
ware range
Setting range,
app
Meaning
Basic flex.
0–200
+/- 20
Level of flexion resistance at the start of flex
ing the knee joint
Gain
0–100
+/- 10
Increase in flexion resistance (starting with
the parameter "
Basic flex.
") when flexing the
knee joint. The knee joint locks at a certain
flexion angle, which depends on the setting
of the parameters "
Basic flex.
" and "
Gain
".
Basic ext.
0–60
+/- 20
Level of extension resistance
Locking angle
0–90
+/- 10
Angle up to which the knee joint can be
extended.
Information:
If this parameter is >0, the knee
joint is locked in a flexed position in the exten
sion direction. To unlock it, unload the pros
thesis and tilt it back for at least 1.5 seconds.
This makes extension of the joint possible
independently of the settings for the paramet
ers "
Basic ext.
" and "
Locking angle
". This
may be necessary to switch to basic mode
using a movement pattern.
Pitch
1000 Hz–4000 Hz 1000 Hz–4000 Hz Pitch of beep signal for confirmation tones
Volume
0–4
0–4
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch
ing MyModes). The "0" setting deactivates
the audible feedback signals. However,
warning signals are still generated if errors
occur.
95