Please ensure that the qualified personnel applying the product is not only authorised for fitting this knee joint, but
also for the connection to the osseointegrated implant system.
3 Application
3.1 Indications for use
The product is to be used
solely
for lower limb exoprosthetic fittings.
3.2 Conditions of use
The product was developed for everyday use and must not be used for unusual activities. These unusual activities
include, for example, extreme sports (free climbing, parachuting, paragliding, etc.).
Permissible ambient conditions are described in the technical data (see Page 103).
The product is intended
exclusively
for use on
one
patient. Use of the product by another person is not approved
by the manufacturer.
Our components perform optimally when paired with appropriate components based upon weight and mobility
grades identifiable by our MOBIS classification information and which have appropriate modular connectors.
Knee joint with attached 2R20 AXON tube adapter
m°
kg
The product is recommended for mobility grade 2 (restricted outdoor walker), mobility grade 3
(unrestricted outdoor walker) and mobility grade 4 (unrestricted outdoor walker with particularly
high demands). Approved for a body weight of up to
150 kg (330 lbs)
.
Knee joint with attached 2R21 AXON tube adapter with torsion
m°
kg
The product is recommended for mobility grade 2 (restricted outdoor walker), mobility grade 3
(unrestricted outdoor walker) and mobility grade 4 (unrestricted outdoor walker with particularly
high demands). Approved for a body weight of up to
125 kg (275 lbs)
.
3.3 Indications
•
For patients with knee disarticulation, transfemoral amputation and hip disarticulation (patients with hip disar
ticulation or hemipelvectomy must be fitted with the 7E10=* Helix
3D
hip joint).
•
For unilateral or bilateral amputation
•
Dysmelia patients with residual limb characteristics corresponding to knee disarticulation, transfemoral ampu
tation or hip disarticulation
•
The patient must fulfil the physical and mental requirements for perceiving visual/acoustic signals and/or mech
anical vibrations.
3.4 Qualification
The product may be fitted only by qualified personnel authorised by Ottobock after completing the corresponding
training.
Only valid for the products 3B1-2/3B1-2=ST
If the product is to be connected to an osseointegrated implant system, the qualified personnel must also be
authorised for the connection to the osseointegrated implant system.