Table 6 – Recommended safe distances between portable and mobile HF telecommunication equipment
and the product 'Genium Bionic Prosthetic System 3B1 / Genium 3B1-2' – for non-life-sustaining devices
and systems
Directives and manufacturer's declaration – Recommended safe distances between portable and mobile
HF telecommunication equipment and the product 'Genium Bionic Prosthetic System 3B1 / Genium
3B1-2'
The product 'Genium Bionic Prosthetic System 3B1 / Genium 3B1-2' is designed for operation in an electromag
netic environment where HF interference is monitored. The customer or user of the product 'Genium Bionic Pros
thetic System 3B1 / Genium 3B1-2' can help prevent electromagnetic interference by complying with the minim
um distances between wearable and mobile HF communication equipment (transmitters) and the product 'Geni
um Bionic Prosthetic System 3B1 / Genium 3B1-2', as below according to the maximum output of the communic
ation equipment.
Nominal output of trans
mitter [W]
Safety distance [m] by transmission frequency
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 0.3 √P
800 MHz to 2.5 GHz
d = 2.3 √P
0.01
N/A
0.03
0.23
0.1
N/A
0.09
0.73
1
N/A
0.3
2.3
10
N/A
0.95
7.3
100
N/A
3
23
For transmitters with no maximum nominal output specified in the table above, the recommended safety distance
d in metres (m) can be determined using the equation in the respective column, where P stands for the maximum
nominal output of the transmitter in watts (W) according to the transmitter manufacturer's information.
Note 1:
The higher frequency range applies at 80 MHz and 800 MHz.
Note 2:
These directives may not be applicable in all cases. The propagation of electromagnetic factors is influ
enced by absorption and reflection by buildings, objects and people.