Short increment indication
The presence of a short increment in intensity is indicated by the little
L
or
R
test ear indicator in the display changing
momentarily to a rectangle,
.
Scoring of responses
•
The patient must use the patient responder to register each increment that has been heard.
•
Manual scoring
If the patient is unable to use the patient responder, but registers an increment heard in another way, you can press
PASS
(
right INT
button) to indicate that the tone was heard.
Pausing the test
•
You can pause the SISI test by pressing
START/STOP
again. During the pause "SISI" will flash in the display. To resume
the test press
START/STOP
once more.
Stopping the test
•
During a pause, you can
stop the test completely
by pressing
ERASE
. Select
Yes
to erase all SISI data and stop the test
completely, or
No
to retain SISI data and resume the test.
Restarting the test
•
If you press
ERASE
when the test is in progress, select
Yes
to erase data. The counter is reset to zero and the test will
start again from the beginning.
Changing the dB level
•
To change the dB level use the
left LEVEL
knob.
Changing the dB increments
•
To change the dB increments use the
Right level
knob (0.25 - 5dB).
If you do so while the test is in progress, the counters are reset to zero and the test starts again from the beginning.
Changing number of increments
•
To change the preset number of SISI dB increments transmitted during a test (10-50) press
SETUP
once and use the
Right level
knob. Then press
SETUP
again to implement the new setting. While the SISI test is in progress, it is not
possible to change the frequency or the number of dB increments.
8.3.1
Masking during SISI
Masking
•
To perform masking, press
MASK ON
or
LOCK
.
When you hold down the button, the masking intensity (instead of % score) and signal type are displayed.
Intensity
•
To change the intensity press and hold the button and turn the
Right level
knob.
Signal type
•
To change the signal type press and hold the button and turn the
left LEVEL
knob.
76
Otometrics - MADSEN Itera II
8 Special tests
Содержание madsen itera ii
Страница 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Страница 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Страница 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Страница 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Страница 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Страница 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Страница 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Страница 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Страница 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...