In order to use the insert phone for masking bone conduction, the EAR-3A must be included as a transducer for your
audiometer. It will be available either as the standard AC transducer, or as an SF transducer. Make sure that it is selec-
ted in
SETUP
,
The test procedure is similar to that described in
Bone conduction with masking - suggested procedure
.
Min IA is frequency specific
These are the Min IA tables for TDH-39 and EAR-3A that you can use to optimize the use of masking.
The numbers are rounded down to the closest 5 dB step.
Min IA TDH-39
Hz
dB
Source
125
35
Katz & Lezynski, (2002)
250
48
Munro & Agnew, BJA (1999)
500
44
Munro & Agnew, BJA (1999)
750
40
N/A - fulfill traditional approach
1000
48
Munro & Agnew, BJA (1999)
1500
40
N/A - fulfill traditional approach
2000
44
Munro & Agnew, BJA (1999)
3000
56
Munro & Agnew, BJA (1999)
4000
50
Munro & Agnew, BJA (1999)
6000
44
Munro & Agnew, BJA (1999)
8000
42
Munro & Agnew, BJA (1999)
Min IA EAR-3A
Hz
dB
Source
125
35
N/A - traditional approach
250
48
Munro & Agnew, BJA (1999)
500
44
Munro & Agnew, BJA (1999)
750
40
N/A - traditional approach
1000
48
Munro & Agnew, BJA (1999)
1500
40
N/A - traditional approach
2000
44
Munro & Agnew, BJA (1999)
3000
56
Munro & Agnew, BJA (1999)
4000
50
Munro & Agnew, BJA (1999)
Otometrics - MADSEN Itera II
65
6 Tone testing
Содержание madsen itera ii
Страница 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Страница 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Страница 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Страница 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Страница 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Страница 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Страница 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Страница 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Страница 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...