19.3
Notes on EMC (Electromagnetic Compatibility)
•
MADSEN Itera II is part of a medical electrical system and is thus subject to special safety precautions. For this reason,
the installation and operating instructions provided in this document should be followed closely.
•
Portable and mobile high-frequency communication devices, such as mobile phones, may interfere with the func-
tioning of MADSEN Itera II.
Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems
MADSEN Itera II is intended for use in the electromagnetic environment specified below. The user of MADSEN Itera II should ensure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
MADSEN Itera II uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
MADSEN Itera II is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Note
: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential envir-
onment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The
user might need to take mitigation measures, such as relocating or re-orienting the equipment.
Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
MADSEN Itera II is intended for use in the electromagnetic environment specified below. The user of MADSEN Itera II should ensure that it is used in such an envir-
onment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with syn-
thetic material, the relative humidity should be at least 30 %.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical loc-
ation in a typical commercial or hospital environment.
U
T
is the AC mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems that are NOT life-supporting
MADSEN Itera II is intended for use in the electromagnetic environment specified below. The user of MADSEN Itera II should ensure that it is used in such an envir-
onment.
124
Otometrics - MADSEN Itera II
19 Technical specifications
Содержание madsen itera ii
Страница 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Страница 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Страница 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Страница 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Страница 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Страница 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Страница 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Страница 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Страница 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...