![otometrics madsen itera ii Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/otometrics/madsen-itera-ii/madsen-itera-ii_reference-manual_1658678021.webp)
•
To adjust the masking intensity use the
Right Level
knob.
To change the default setting, select
SETUP
,
.
OFF
No masking
ON
To adjust the masking intensity independently of the signal level use the
Right level
knob.
LOCK
Makes the masking level follow the stimulus signal level. You can adjust the difference between the stimulus
signal and the masking signal by using the
Right level
knob.
MASK shortcut functions
Masking transducer selection
When masking is already selected and you press the same masking button once
more, the masking transducer, the masking signal type and the name of the trans-
ducer, and (if applicable) the serial no. appear in the display.
The transducer button flashes and you can then select a different masking trans-
ducer by pressing one of the three
TRANSD
buttons
AC
,
BC
or
SF
.
Masking noise in Speech
Select the masking ear: press and hold
MASK ON
and press
L <--> R
to toggle
between left, right and both ears.
3.2.12
dB steps
Choose between the following attenuator steps:
1 dB
2.5 dB
5 dB
If the indicator light on the selected button flashes, this indicates that the display has been set up to show
SPL values (
SETUP
,
) instead of HL.
3.2.13
Microphone
The internal microphone for Talk Over and Speech testing.
Use this microphone as an alternative to the goose neck microphone or the boom microphone on the mon-
itor headset.
Otometrics - MADSEN Itera II
21
3 Overview of MADSEN Itera II
Содержание madsen itera ii
Страница 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Страница 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Страница 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Страница 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Страница 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Страница 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Страница 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Страница 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Страница 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...