Otocheck ABR
User Manual
71
Configuration
When to use each PC:
PC1 is ideal for performing an ABR test on adult subjects and is the
default setting for ABRcustom mode. It is useful for training, demonstration
and clinical investigations.
PC2 is designed for neonate screening and is the default setting for the
ABRscreen mode.
PC3 and PC4 may give higher specificity in challenging noise
environments, potentially passing healthy ears securely even when
interference levels are high. Both are designed for neonate screening but
PC4 can also be used with adult ears.
Ear Cups
This setting allows the use of ear cups and it can be set in three different
ways :
OFF – Will always go into test assuming ear cups are not being used.
AUTO – Will detect whether ear cups are being used and prompt the user
before test if ear cups are being used or not.
ON – Will always go into test assuming ear cups are being used.
ABR mains
ABR mains sets the ABR test configuration for optimum test recording in
mains (powerline) noise. The mains frequency varies in different countries.
Set 50Hz if the mains in your region is 50Hz and 60Hz if your region is
60Hz. Select Not Set if you are unsure of the mains frequency in your
region.
Setting ABR mains from the main menu
If ABR mains is
Not Set
then you will be prompted to set it when an ABR
test is selected from the main menu: