CHAPTER SIXTEEN
Care of the Otocheck ABR
114
Care of the Otoport
Before fitting each tip, ensure the sound tubes are carefully examined for
any sign of debris that may have entered them. Replace any part of the
probe ear piece as necessary. (See chapter
15
Probe
care
for details)
Ensure your hands are cleaned thoroughly between each patient tested.
Clean the Otocheck each day before a testing session, or according
to local requirements. Ensure the Otocheck is cleaned if it becomes
contaminated. Clean surfaces of the Otocheck with an alcohol based
sterile wipe or cloth and antiseptic fluid. Dry the Otocheck with tissue
immediately afterwards. Do not allow liquid to enter the instrument and
do not immerse in fluid. Do not allow liquid to come into contact with the
connection sockets. Do no poke any materials into either the probe or
charger/pc cable sockets.
If additional hygienic protection is required, use the Otocheck in an
infection control sleeve. This can also be cleaned with a sterile wipe
or cloth with antiseptic fluid. The sleeves are disposable, so should be
replaced weekly or after approximately every 50 tests.