126 CHAPTER EIGHTEEN
Equipment troubleshooting
Equipment troubleshooting
18.5
Hardware fault messages
The Otocheck performs a series of hardware tests when it is first turned
on. In the event of a fault being detected the following message will be
displayed:
The error code number displayed indicates the type of error detected.
You should make a note of this error number. The Otocheck should then
be turned on and off a number of times to ensure that the error doesn’t
reoccur.
If you receive the hardware fault message again, contact your distributor or
Otodynamics for support.
Error 2
The above message indicates that excessive noise was detected during
start up. The noise may have been detected through the probe, if it was
connected. Noise detected through the probe does not indicate a fault. If
this message is displayed, turn the unit off, disconnect the probe and then
turn it back on. If the message is consistently displayed with the probe
disconnected, then contact Otodynamics.