19
(V) Save All
•
Clicking this button will manually save all the measurement results that are currently displayed in
the graph.
The Ossila Potentiostat includes a Test Cell Chip, shown in
Figure 9.10
, which can be used to check that your
Potentiostat is operating correctly. It simulates an electrochemical cell by providing a response which differs
depending on the direction of the potential scan.
9.3 Test Cell Chip
Figure 9.10.
Ossila Test Cell Chip.
Figure 9.11.
Settings for measuring the Test Cell Chip.
9.3.1 Taking a Measurement
1. Plug in and turn on the Potentiostat.
2. Use the cell connection cable to connect the Potentiostat to the same colour connectors on the
Test Cell Chip.
I. Red connector to WE1, 2, 3, or 4.
II. Black connector to CE.
III. Blue connector to RE.
3. Start the Ossila Cyclic Voltammetry software. The window shown in
Figure 9.2
will open.
4. In the software enter the settings shown in
Figure 9.11
.
5. Click the ‘Measure’ button.