15
9.2.2 Saving and Loading Settings
Figure 9.4.
Controls for saving and loading settings profiles.
(II) Current Range
•
Select the range of currents to be used for the measurement or automatic range selection.
I. This defines the upper limit, accuracy, and resolution of the current measurements that can be
performed by the system.
II. Automatic range selection will start on the lowest current range and automatically switch to
higher ranges if the current increases above the maximum for a range.
(III) Start Potential (V)
•
The potential in volts at which the measurement starts.
(IV) Potential Vertex 1 (V)
•
The first potential in volts at which the scan changes direction.
(V) Potential Vertex 2 (V)
•
The second potential in volts at which the scan changes direction.
(VI) Scan Rate (mV/s)
•
The rate at which the potential will be changed during the scan, measured in millivolts per second.
(VII) Cycles
•
The number of times the scan will be repeated.
(I) Save Settings
•
Saves the current settings as a profile that can be loaded quickly for use at another time.
•
When clicked, you will be prompted to name the settings profile.
I. If the name is already in use, you will be asked if you wish to overwrite the previous profile.
II. The name cannot contain the characters: \ / : * ? “ < > |
III. You can change the default settings by choosing the name ‘Default’.
•
The settings profile will be added to the drop-down box using the given profile name.