Four-Point Probe System
User Manual
13
Ossila Limited © 2022
9.2.1
Basic Settings
Figure 9.2.
Basic settings.
(I)
System Address
•
Select the COM port or IP address of the connected unit you intend to use (USB and
Ethernet connection respectively).
I.
This box will be populated automatically with the addresses of any units connected
to the computer or network.
(II)
Geometry
•
Select the geometry of the sample being measured.
I.
This is required to calculate the geometrical correction factor for the current
sample.
II.
If the shape of the sample is irregular, consider whether it is closer to rectangular
or circular and then estimate what size of that shape could fit within the sample.
(III)
Long Side (Rectangular Sample)
•
Sets the length of the long side of the sample in mm (if the sample is rectangular).
I.
This is required for calculating the appropriate geometrical correction factor.
(IV)
Short Side (Rectangular Sample)
•
Sets the length of the short side of the sample in mm (if the sample is rectangular).
I.
This is required for calculating the appropriate geometrical correction factor.
(V)
Diameter (Circular Sample)
•
Sets the diameter of the sample in mm (if the sample is circular).
I.
This is required for calculating the appropriate geometrical correction factor.