Four-Point Probe System
User Manual
20
Ossila Limited © 2022
11.2
Error Messages and Warnings
Message
Description
Current limit exceeded
The measured current is greater than the set current limit.
Voltage limit exceeded
The output voltage is greater than the set voltage limit.
Unable to measure sample,
resistance outside of
measurable range
The system is unable to pass a current through the sample
under test.
Long edge must be able to fit
all probes
The long edge of a rectangular sample is shorter than 3
times the probes spacing.
Diameter must be able to fit all
probes
The diameter of a circular samples is shorter than 3 times
the probes spacing.
Short edge cannot be longer
than the long edge
The short edge is greater than the long edge for a
rectangular sample.
Sample removed, stopping
measurement
The system has detected that the sample has been removed.
Error communicating with
system
The software is unable to connect to the system.
Unable to open file:
permission was denied
The software does not have the necessary permissions to
access the given file path, or the file is already open in other
software.
Unable to save file: permission
was denied
The software does not have the necessary permissions to
access the given file path, or the file is already open in other
software.