10.
Dip Coater Troubleshooting
Problem
Possible cause
Action
No power / display
The power switch on the unit is in the
OFF position
Check the connection and ensure the
power is turned ON
The power supply may not be
connected properly
Ensure the unit is firmly plugged in to the
power supply, and the plug is firmly
connected to both the adapter and the
working power socket
The fuse on the rear panel has blown
Ensure the unit is unplugged
Check the fuse on the rear panel. If it has
blown, replace with a suitably rated 1A
slow blow fuse
The power supply adapter has a fault
Contact Ossila for a replacement power
supply adapter
No obvious cause
If all the above causes have been
considered, there may be a fault on the
board.
Please
contact
Ossila
for
information
Crash Warning
The safety switch in the crash switch
housing has not been returned to its
ON state.
Ensure that the arm position is returned to
its initial position and that no obstruction
to the arms movement remains
The safety switch in the crash switch
housing has developed a fault
If the above has not resolved the issue,
please
contact
Ossila
for
more
information
System Error
The secondary safety switch has
accidentally been triggered
Turn the unit off and on again to see if the
warning remains
Internal fault in the secondary safety
switch
If all the above has been considered there
may be a fault with the secondary safety
switch;
contact
Ossila
for
more
information