Troubleshooting 

(cont.)

 

 

 

 

 

 

Reference Sensor’s internal temperature 

exceeds limit of 70°C.  Wait for sensor to 

cool  down, then continue. 

Repeat current step. If error  persists, 
reference sensor  battery may need to 

be replaced. 

Replace reference sensor battery. 

 

 

 

 

 

Confirm that all connections are secure. 

 
 
 

 

 

Choose ‘Yes’ or ‘No’ depending on whether 

pelvic base or femur tack has moved during 

surgery. If ‘Yes’ is selected, leg length and 

offset  measurements will be disabled. 

Remove reference sensor and clean probe 
markings, clean reference sensor window 

located on  bottom side and re-attach 

reference sensor.  If  problem persists, 

restart reference sensor by  removing and 

replacing reference sensor battery. 

 

 

There is a major system fault and 

navigation  unit must be turned off and 

replaced. Please  recover faulty unit and 

return to manufacturer  for analysis. 

Remove femur tack and screws 

from the  wound prior to closure. 

 
 
 

 

 

A software error has occurred. Power unit 

off  and on again only one time.  If System 

fault reoccurs after re-starting, power unit 
off and replace it.  Contact OrthAlign and 

return unit for  further analysis. 

Software Error

 

 

Press and hold left 

and up buttons to 

turn unit off

 

Then restart

 

Warning:

 

Has Femur Tack or 
Pelvic Base moved?

 

Yes

 

No

 

Warning:

 

Remove Femur Tack and Pelvic 

Base

 

Warning

 

Reference  Sensor 

too hot 

89.0º C

 

Press center button 

to retry

 

Содержание HipAlign

Страница 1: ...HipAlign Surgical Technique Manual Total Hip Arthroplasty Posterior Approach...

Страница 2: ...Base 9 Attaching Sensors and Registration of Landmarks 10 Attaching Sensors 10 Aligning Jig 11 Adjusting the Registration Jig 11 How to Register Landmarks 11 Registrations 12 Options Menu 13 Prepare...

Страница 3: ...ut 1mW emitted wavelength 635nm duration 0 25 seconds continuous beam divergence 10 degrees per IEC 60825 1 2014 05 This information can be found etched onto the back surface of the laser module Do no...

Страница 4: ...vice such as a peg board or vise type patient positioner WARNING Correct patient positioning is essential for accurate navigation The system calculates cup angles and changes in leg length and joint o...

Страница 5: ...number matches the one marked on reference sensor being used Otherwise press left button to let navigation unit find another reference sensor Repeat until correct reference sensor serial number is di...

Страница 6: ...k that navigation unit and reference sensor are correctly attached to the bracket and press left button three times to repeat from first calibration step Step 4 STEP 4 Angle unit away from you so that...

Страница 7: ...Anterior 403306 Pelvic Bracket Assembly Short 403307 Straight Cup Impactor 403338 Reference Sensor 5 Hip RS5H 405001 or 403087 06 Threaded Fixation Pin 4 0mm x 110mm with 3 2mm Proximal End 403317 11...

Страница 8: ...Pre Op Selections Step 10 Select Hip Select operative hip Step 12 Input Abduction Input target cup abduction angle This is the Radiographic Inclination angle defined as the coronal plane projection of...

Страница 9: ...elvic base as a guide for spacing between the pins Slide the Pelvic Base over both pins to the level of the skin and use the 3 5mm Hex Driver to tighten the two bottom screws to secure Base WARNING Th...

Страница 10: ...reference sensor to probe by aligning arrows on sensor and probe TIP Attach pelvic bracket with navigation unit attached to coupler on pelvic base while pushing on pelvic bracket lever TIP Attach prob...

Страница 11: ...gh plate is aligned with original marking on skin or if unable to return it to original position Adjusting the Registration Jig TIP To adjust the laser alignment on the leg loosen pelvic base locking...

Страница 12: ...tration marker Open laser shutter place the vertical target on the thigh plate and align the leg so that the laser projects onto the lower part of the target Trace the crosshair pattern with a marking...

Страница 13: ...hread the adapter into the acetabular shell The release lever can be used to release the adapter from the impactor Offset Impactor Snap the adapter into the impactor Thread the adapter into the acetab...

Страница 14: ...ated according to their radiographic definitions NOTE The navigation unit proceeds to the next step if 30 second timer expires center button is pressed reference sensor detects an impact strong enough...

Страница 15: ...the laser Verify Homepoint Step V1 Attach Sensors Attach probe and probe mount to bracket Attach bracket to pelvic base Step V2 Verify Homepoint Place tip of probe into hole of Pelvic Base Press cente...

Страница 16: ...off navigation unit by pressing left and up arrow buttons until unit is turned off Additionally remove unit from pelvic bracket by pushing down on top lever NOTE The unit will power off after 90 minut...

Страница 17: ...ence Sensor 5 Hip RS5H 1 403317 110 Threaded Fixation Pin 4 0mm x 110mm with 3 2mm Proximal End 3 403376 Thigh Plate Slotted 1 403377 Vertical Target Magnet 1 Additional Instruments 403402 Femur Regis...

Страница 18: ...lization Position Instruments in Instrument Tray as shown below for sterilization Refer to the Instructions for Use for details on cleaning and sterilization parameters Vertical Laser Target Shell Ada...

Страница 19: ...fic light turns green The reference sensor was moved too quickly Make sure reference sensor is moved smoothly from one position to the next without rapid rotations Place reference sensor in docked pos...

Страница 20: ...reference sensor If problem persists restart reference sensor by removing and replacing reference sensor battery There is a major system fault and navigation unit must be turned off and replaced Pleas...

Страница 21: ...effective surgical navigation products for precise alignment For more information about the OrthAlign Plus System please contact us at 949 715 2424 or info orthalign com 120 Columbia Suite 500 Aliso V...

Отзывы: