12 

 

Registrations

 

 

Register Home

 

 

Place tip of probe into hole of the pelvic base and press center button. 

 
 
 

 

 

 
 
 
 
 
 

WARNING:

 

Register Table

 

 

Position operating table horizontally. 

 

Verify the sagittal plane of pelvis is level. 

 

Move probe, with reference sensor attached, to align parallel long axis of body 
with the tip pointing towards the feet. 

 

Press the center button to register. 

 

Correct patient positioning is essential for accurate navigation. The system calculates cup angles based on the assumption 
that the pelvis is accurately positioned during table registration.

 

 

Table registration must be completed prior to dislocating hip, as torque applied during dislocation can move pelvis away 
from the correct initial alignment.

 

 

Pelvic Base cannot be moved after this point.

 

 

Pelvic Bracket angle to the base cannot be adjusted after this point.

 

 

If fixation pins become loose, navigation must be aborted, since table registration is no longer valid.

 

 
 
 

Registering Femur

 

 

Make the standard incision and identify a point along greater 
trochanter to insert a registration marker for femoral 
registration. Insert femur registration marker. 

 

Open laser shutter, place the vertical target on the thigh plate and 
align the leg so that the laser projects onto  the lower part of the 
target. 

 

Trace the crosshair pattern with a marking pen. 

 

While laser projection is aligned with laser marking, hold the 
probe tip in the center hole of the marker and press the  center 
button to register the femur.  

 

Close shutter on the laser and remove vertical laser target. 

TIPS

:

 

1.

 

When marking the vertical target and registering the femur 
registration marker, position  the thigh in neutral flexion, 
abduction and rotation.

 

2.

 

Stack legs vertically atop each other in a stable position which 
will be easy to return to prior to registration.

 

 

WARNING:

 

 

If laser module moves, leg length must be aborted since leg alignment cannot be replicated.

 

 

Discontinue use if unable to verify thigh plate is aligned with original marking on skin or if unable to return it to 
original position.

 

Prepare Acetabulum

 

NOTE

: Once registration is complete, proceed to dislocate the hip, resect the femoral head, and prepare the

 

acetabulum as per the implant manufacturer’s technique.

 

H4

 

H3

 

Register Home

 

 

Place probe tip in hole 

of pelvic base

 

Keep pelvic base locked

 

throughout procedure

 

Register Table

 

Verify sagittal plane of pelvis is level and 

align probe with coronal plane

 

Содержание HipAlign

Страница 1: ...HipAlign Surgical Technique Manual Total Hip Arthroplasty Posterior Approach...

Страница 2: ...Base 9 Attaching Sensors and Registration of Landmarks 10 Attaching Sensors 10 Aligning Jig 11 Adjusting the Registration Jig 11 How to Register Landmarks 11 Registrations 12 Options Menu 13 Prepare...

Страница 3: ...ut 1mW emitted wavelength 635nm duration 0 25 seconds continuous beam divergence 10 degrees per IEC 60825 1 2014 05 This information can be found etched onto the back surface of the laser module Do no...

Страница 4: ...vice such as a peg board or vise type patient positioner WARNING Correct patient positioning is essential for accurate navigation The system calculates cup angles and changes in leg length and joint o...

Страница 5: ...number matches the one marked on reference sensor being used Otherwise press left button to let navigation unit find another reference sensor Repeat until correct reference sensor serial number is di...

Страница 6: ...k that navigation unit and reference sensor are correctly attached to the bracket and press left button three times to repeat from first calibration step Step 4 STEP 4 Angle unit away from you so that...

Страница 7: ...Anterior 403306 Pelvic Bracket Assembly Short 403307 Straight Cup Impactor 403338 Reference Sensor 5 Hip RS5H 405001 or 403087 06 Threaded Fixation Pin 4 0mm x 110mm with 3 2mm Proximal End 403317 11...

Страница 8: ...Pre Op Selections Step 10 Select Hip Select operative hip Step 12 Input Abduction Input target cup abduction angle This is the Radiographic Inclination angle defined as the coronal plane projection of...

Страница 9: ...elvic base as a guide for spacing between the pins Slide the Pelvic Base over both pins to the level of the skin and use the 3 5mm Hex Driver to tighten the two bottom screws to secure Base WARNING Th...

Страница 10: ...reference sensor to probe by aligning arrows on sensor and probe TIP Attach pelvic bracket with navigation unit attached to coupler on pelvic base while pushing on pelvic bracket lever TIP Attach prob...

Страница 11: ...gh plate is aligned with original marking on skin or if unable to return it to original position Adjusting the Registration Jig TIP To adjust the laser alignment on the leg loosen pelvic base locking...

Страница 12: ...tration marker Open laser shutter place the vertical target on the thigh plate and align the leg so that the laser projects onto the lower part of the target Trace the crosshair pattern with a marking...

Страница 13: ...hread the adapter into the acetabular shell The release lever can be used to release the adapter from the impactor Offset Impactor Snap the adapter into the impactor Thread the adapter into the acetab...

Страница 14: ...ated according to their radiographic definitions NOTE The navigation unit proceeds to the next step if 30 second timer expires center button is pressed reference sensor detects an impact strong enough...

Страница 15: ...the laser Verify Homepoint Step V1 Attach Sensors Attach probe and probe mount to bracket Attach bracket to pelvic base Step V2 Verify Homepoint Place tip of probe into hole of Pelvic Base Press cente...

Страница 16: ...off navigation unit by pressing left and up arrow buttons until unit is turned off Additionally remove unit from pelvic bracket by pushing down on top lever NOTE The unit will power off after 90 minut...

Страница 17: ...ence Sensor 5 Hip RS5H 1 403317 110 Threaded Fixation Pin 4 0mm x 110mm with 3 2mm Proximal End 3 403376 Thigh Plate Slotted 1 403377 Vertical Target Magnet 1 Additional Instruments 403402 Femur Regis...

Страница 18: ...lization Position Instruments in Instrument Tray as shown below for sterilization Refer to the Instructions for Use for details on cleaning and sterilization parameters Vertical Laser Target Shell Ada...

Страница 19: ...fic light turns green The reference sensor was moved too quickly Make sure reference sensor is moved smoothly from one position to the next without rapid rotations Place reference sensor in docked pos...

Страница 20: ...reference sensor If problem persists restart reference sensor by removing and replacing reference sensor battery There is a major system fault and navigation unit must be turned off and replaced Pleas...

Страница 21: ...effective surgical navigation products for precise alignment For more information about the OrthAlign Plus System please contact us at 949 715 2424 or info orthalign com 120 Columbia Suite 500 Aliso V...

Отзывы: