background image

15 

 

Check Leg Length

 

 

Step F5: Attach Sensors

 

 

Attach probe with reference sensor to jig 

 
 
 

 

Step F5A: Position Femur

 

 

Position the magnetic vertical target on the distal thigh 

 

Open shutter on laser 

 

Re-align leg so that the laser projection aligns with the marking on the vertical target. 

NOTE: 

Ensure laser lens is clear. 

 

WARNING: 

 

 Discontinue use of laser if crosshair pattern is obscured.

 

 

Discontinue use if unable to verify thigh plate is aligned with original marking on skin 
or if unable to return it to original position.

 

Step F6: Register Femur

 

 

Hold the probe tip in the center hole of the screw and press the center button to register the femur. 

 

Remove vertical laser target. 

 

 

Step F9: Register Femur

 

 

Leg Length and offset are displayed. 

 

Close the shutter on  the laser. 

 

 

 

 

Verify Homepoint

 

 
 

Step V1: Attach Sensors

 

 

 

Attach probe and probe mount to bracket. Attach bracket to pelvic 
base. 

 
 

Step V2: Verify Homepoint

 

 

 

Place tip of probe into hole of Pelvic Base. 

 

Press center button to register Homepoint. 

 

Step V3: Homepoint Accuracy

 

 

 

View change in Homepoint which is change in probe tip position between 
original Homepoint and current Homepoint registrations. 

 

 

 

NOTE: 

The number displayed may not be zero due to mechanical play and reference sensor noise. If displayed number is 

greater than 3mm, it is suggested to verify all reference sensor attachments and jig connections are correct and secure 
before repeating.  Repeat registrations if accuracy or previous registrations is in question.

 

V1 

Attach Sensors

 

 

Attach probe to 

probe mount

 

Attach probe mount

 

to jig

 

Attach jig to pelvic base

 

V2 

Verify Homepoint

 

 

Place probe tip in

 

hole of pelvic base

 

 

Keep  pelvic  base 

locked  throughout 

procedure

 

V3 

Homepoint Accuracy

 

 

Homepoint 

:   

mm

 

 

Содержание HipAlign

Страница 1: ...HipAlign Surgical Technique Manual Total Hip Arthroplasty Posterior Approach...

Страница 2: ...Base 9 Attaching Sensors and Registration of Landmarks 10 Attaching Sensors 10 Aligning Jig 11 Adjusting the Registration Jig 11 How to Register Landmarks 11 Registrations 12 Options Menu 13 Prepare...

Страница 3: ...ut 1mW emitted wavelength 635nm duration 0 25 seconds continuous beam divergence 10 degrees per IEC 60825 1 2014 05 This information can be found etched onto the back surface of the laser module Do no...

Страница 4: ...vice such as a peg board or vise type patient positioner WARNING Correct patient positioning is essential for accurate navigation The system calculates cup angles and changes in leg length and joint o...

Страница 5: ...number matches the one marked on reference sensor being used Otherwise press left button to let navigation unit find another reference sensor Repeat until correct reference sensor serial number is di...

Страница 6: ...k that navigation unit and reference sensor are correctly attached to the bracket and press left button three times to repeat from first calibration step Step 4 STEP 4 Angle unit away from you so that...

Страница 7: ...Anterior 403306 Pelvic Bracket Assembly Short 403307 Straight Cup Impactor 403338 Reference Sensor 5 Hip RS5H 405001 or 403087 06 Threaded Fixation Pin 4 0mm x 110mm with 3 2mm Proximal End 403317 11...

Страница 8: ...Pre Op Selections Step 10 Select Hip Select operative hip Step 12 Input Abduction Input target cup abduction angle This is the Radiographic Inclination angle defined as the coronal plane projection of...

Страница 9: ...elvic base as a guide for spacing between the pins Slide the Pelvic Base over both pins to the level of the skin and use the 3 5mm Hex Driver to tighten the two bottom screws to secure Base WARNING Th...

Страница 10: ...reference sensor to probe by aligning arrows on sensor and probe TIP Attach pelvic bracket with navigation unit attached to coupler on pelvic base while pushing on pelvic bracket lever TIP Attach prob...

Страница 11: ...gh plate is aligned with original marking on skin or if unable to return it to original position Adjusting the Registration Jig TIP To adjust the laser alignment on the leg loosen pelvic base locking...

Страница 12: ...tration marker Open laser shutter place the vertical target on the thigh plate and align the leg so that the laser projects onto the lower part of the target Trace the crosshair pattern with a marking...

Страница 13: ...hread the adapter into the acetabular shell The release lever can be used to release the adapter from the impactor Offset Impactor Snap the adapter into the impactor Thread the adapter into the acetab...

Страница 14: ...ated according to their radiographic definitions NOTE The navigation unit proceeds to the next step if 30 second timer expires center button is pressed reference sensor detects an impact strong enough...

Страница 15: ...the laser Verify Homepoint Step V1 Attach Sensors Attach probe and probe mount to bracket Attach bracket to pelvic base Step V2 Verify Homepoint Place tip of probe into hole of Pelvic Base Press cente...

Страница 16: ...off navigation unit by pressing left and up arrow buttons until unit is turned off Additionally remove unit from pelvic bracket by pushing down on top lever NOTE The unit will power off after 90 minut...

Страница 17: ...ence Sensor 5 Hip RS5H 1 403317 110 Threaded Fixation Pin 4 0mm x 110mm with 3 2mm Proximal End 3 403376 Thigh Plate Slotted 1 403377 Vertical Target Magnet 1 Additional Instruments 403402 Femur Regis...

Страница 18: ...lization Position Instruments in Instrument Tray as shown below for sterilization Refer to the Instructions for Use for details on cleaning and sterilization parameters Vertical Laser Target Shell Ada...

Страница 19: ...fic light turns green The reference sensor was moved too quickly Make sure reference sensor is moved smoothly from one position to the next without rapid rotations Place reference sensor in docked pos...

Страница 20: ...reference sensor If problem persists restart reference sensor by removing and replacing reference sensor battery There is a major system fault and navigation unit must be turned off and replaced Pleas...

Страница 21: ...effective surgical navigation products for precise alignment For more information about the OrthAlign Plus System please contact us at 949 715 2424 or info orthalign com 120 Columbia Suite 500 Aliso V...

Отзывы: