Back-Table Set Up

 

 

 

Select Joint

 

Use center button to select Hip. 

 
 
 
 

 

Select Approach

 

Use center button to select approach. 

 
 
 
 
 

 

STEP 1: Install Sensor Battery

 

 

Insert battery into reference sensor and close cover. 

 

TIP: 

Reference sensor LED will slowly flash yellow to indicate 

battery has been inserted correctly. LED will momentarily flash 
green and navigation unit will move to next screen when reference 
sensor is found.

 

 

 

STEP 2: Confirm Sensor ID

 

 

Press center button if displayed number matches the one marked on reference sensor being 
used. Otherwise, press left button to let navigation unit find another reference sensor. Repeat 
until correct reference sensor serial number is displayed. 

 
 

STEP 3: Attach to Pelvic Jig

 

 

Slide the navigation unit onto jig toward lever. Navigation unit should be firmly secured in 
place on the jig. Attach reference sensor to coupler at other end of the pelvic jig as shown. 

 

NOTE: 

Navigation unit must be secure on the pelvic jig before proceeding.

 

 

 

 

S1

 

Select Joint

 

Knee

 

Hip

 

S2

 

Select Approach

 

Posterior

 

Anterior

 

1

 

Install Sensor Battery

 

Seeking  Sensor…

 

2

 

Confirm Sensor ID

 

Confirm

 

####

 

Use Count

 

22

 

No

 

Yes

 

Attach to Pelvic Jig

 

<

 

<

 

<

 

<

 

Содержание HipAlign

Страница 1: ...HipAlign Surgical Technique Manual Total Hip Arthroplasty Posterior Approach...

Страница 2: ...Base 9 Attaching Sensors and Registration of Landmarks 10 Attaching Sensors 10 Aligning Jig 11 Adjusting the Registration Jig 11 How to Register Landmarks 11 Registrations 12 Options Menu 13 Prepare...

Страница 3: ...ut 1mW emitted wavelength 635nm duration 0 25 seconds continuous beam divergence 10 degrees per IEC 60825 1 2014 05 This information can be found etched onto the back surface of the laser module Do no...

Страница 4: ...vice such as a peg board or vise type patient positioner WARNING Correct patient positioning is essential for accurate navigation The system calculates cup angles and changes in leg length and joint o...

Страница 5: ...number matches the one marked on reference sensor being used Otherwise press left button to let navigation unit find another reference sensor Repeat until correct reference sensor serial number is di...

Страница 6: ...k that navigation unit and reference sensor are correctly attached to the bracket and press left button three times to repeat from first calibration step Step 4 STEP 4 Angle unit away from you so that...

Страница 7: ...Anterior 403306 Pelvic Bracket Assembly Short 403307 Straight Cup Impactor 403338 Reference Sensor 5 Hip RS5H 405001 or 403087 06 Threaded Fixation Pin 4 0mm x 110mm with 3 2mm Proximal End 403317 11...

Страница 8: ...Pre Op Selections Step 10 Select Hip Select operative hip Step 12 Input Abduction Input target cup abduction angle This is the Radiographic Inclination angle defined as the coronal plane projection of...

Страница 9: ...elvic base as a guide for spacing between the pins Slide the Pelvic Base over both pins to the level of the skin and use the 3 5mm Hex Driver to tighten the two bottom screws to secure Base WARNING Th...

Страница 10: ...reference sensor to probe by aligning arrows on sensor and probe TIP Attach pelvic bracket with navigation unit attached to coupler on pelvic base while pushing on pelvic bracket lever TIP Attach prob...

Страница 11: ...gh plate is aligned with original marking on skin or if unable to return it to original position Adjusting the Registration Jig TIP To adjust the laser alignment on the leg loosen pelvic base locking...

Страница 12: ...tration marker Open laser shutter place the vertical target on the thigh plate and align the leg so that the laser projects onto the lower part of the target Trace the crosshair pattern with a marking...

Страница 13: ...hread the adapter into the acetabular shell The release lever can be used to release the adapter from the impactor Offset Impactor Snap the adapter into the impactor Thread the adapter into the acetab...

Страница 14: ...ated according to their radiographic definitions NOTE The navigation unit proceeds to the next step if 30 second timer expires center button is pressed reference sensor detects an impact strong enough...

Страница 15: ...the laser Verify Homepoint Step V1 Attach Sensors Attach probe and probe mount to bracket Attach bracket to pelvic base Step V2 Verify Homepoint Place tip of probe into hole of Pelvic Base Press cente...

Страница 16: ...off navigation unit by pressing left and up arrow buttons until unit is turned off Additionally remove unit from pelvic bracket by pushing down on top lever NOTE The unit will power off after 90 minut...

Страница 17: ...ence Sensor 5 Hip RS5H 1 403317 110 Threaded Fixation Pin 4 0mm x 110mm with 3 2mm Proximal End 3 403376 Thigh Plate Slotted 1 403377 Vertical Target Magnet 1 Additional Instruments 403402 Femur Regis...

Страница 18: ...lization Position Instruments in Instrument Tray as shown below for sterilization Refer to the Instructions for Use for details on cleaning and sterilization parameters Vertical Laser Target Shell Ada...

Страница 19: ...fic light turns green The reference sensor was moved too quickly Make sure reference sensor is moved smoothly from one position to the next without rapid rotations Place reference sensor in docked pos...

Страница 20: ...reference sensor If problem persists restart reference sensor by removing and replacing reference sensor battery There is a major system fault and navigation unit must be turned off and replaced Pleas...

Страница 21: ...effective surgical navigation products for precise alignment For more information about the OrthAlign Plus System please contact us at 949 715 2424 or info orthalign com 120 Columbia Suite 500 Aliso V...

Отзывы: