![ORTEC Detective-DX-100T Скачать руководство пользователя страница 95](http://html1.mh-extra.com/html/ortec/detective-dx-100t/detective-dx-100t_user-manual_1654401095.webp)
931040P / 1014
8. USING THE DETECTIVE-EX WITH ORTEC SPECTRUM ANALYSIS PROGRAMS
15
Patent pending.
85
Figure 86. Detective-EX Amplifier 2 Tab.
Figure 87. Detective-EX Amplifier PRO Tab.
8.3.2. Amplifier 2
Figure 86 shows the Amplifier 2 tab,
which displays the fixed Rise Time
and Flattop Width settings. You can
also use the InSight Virtual Oscillo-
scope™ mode to view the digital signal
processor’s actual sampled waveform
on a reference graticule; however, the
shaping settings cannot be adjusted.
8.3.3. Amplifier PRO
This tab (Fig. 87) displays the control
for the Low Frequency Rejector (LFR)
15
feature, a digital filter that corrects the
pulse output signal for cooler-induced
microphonics). It cannot be disabled.