ORTEC
®
Detective-EX/DX
®
-100T Portable HPGe Spectrometer — User’s Manual
931040P / 1014
82
Figure 84. Connected to Com-
puter Via CONTROL Port.
The final step is to run the MCB Configuration pro-
gram according to the instructions in the MAESTRO
User’s Manual to build Master Instrument List of all
ORTEC MCBs accessible to the computer.
To monitor the battery time remaining from within
MAESTRO, go to the Status tab under Acquire/MCB
Properties....
To return the Detective-EX to standalone operation,
simply disconnect from the computer. In a few seconds,
the touchscreen will become active in the normal stand-
alone mode. If, after unplugging the USB cable, the
instrument does not return to the Detective-EX inter-
face, reconnect the cable, wait 12–15 seconds, and dis-
connect. Repeat as needed until the display unlocks.
NOTE
MAESTRO and other ORTEC spectroscopy
applications have a Lock/Unlock Detector command to prevent unauthorized users
from changing detector settings, starting and stopping data acquisition, and clearing
the detector memory. This locking command functions separately from the Detec-
tive-EX password. If the Detective-EX is password-locked within MAESTRO, you
can still use it in the field without knowing the password; however, you cannot recon-
nect it to a computer and erase its memory. In addition, the next time you use it in
MCA Mode with MAESTRO, it will still be locked. Also note that if the Detective-
EX’s display is locked, it cannot communicate with a computer (Section ?).
8.2. Quantitative Analysis of Detective-EX Spectra in
GammaVision and ISOTOPIC
The Detective-EX can be used to collect data for quantitative analysis, either as a standalone
field instrument or a benchtop MCB connected to a computer. In both usage modes, the data are
stored as
.SPC
-format spectrum files. The
.SPC
files can then be quantitatively analyzed on a
computer using GammaVision or ISOTOPIC.
Note, however, that the Detective-EX is not designed to store efficiency information in the
.SPC
files it creates. Therefore, for quantitative determinations, special attention is needed to prepare
the calibration (
.CLB
) file so that GammaVision or ISOTOPIC will add the efficiency informa-
tion during processing.
Содержание Detective-DX-100T
Страница 4: ...iv Intentionally blank ...
Страница 10: ...x ...
Страница 38: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 28 Intentionally blank ...
Страница 50: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 40 Intentionally blank ...
Страница 84: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 74 Intentionally blank ...
Страница 108: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 98 Intentionally blank ...
Страница 112: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 102 Intentionally blank ...