To check these settings:
!
On the third screen of the Advanced menu (Section 5.3.2), tap Exit, then confirm that
you wish to end the Detective-EX program.
!
On the Windows desktop, tap the Start square in the upper left corner of the screen, then
tap Settings.
!
In the Settings dialog, tap the System tab, then select Regional Settings.
!
On the Region tab, select English (United States) from the listbox.
!
On the Number tab, choose the period ( . ) for the Decimal symbol and the comma ( , )
for the Digit grouping symbol.
!
On the Time tab, select h:mm:ss tt as the Time style and the full colon ( : ) as the Time
separator.
!
On the Date tab, use M/d/yy as the Short date, the forward slash ( / ) as the Date
separator, and the Long date format dddd, MMMM dd, yyyy.
!
Tap the upper right OK circle, then the upper right X circle to return to the Windows
desktop.
!
Restart the Detective-EX as described in Section 2.8.2.
For further assistance, contact your ORTEC representative or our Global Service Center.
Содержание Detective-DX-100T
Страница 4: ...iv Intentionally blank ...
Страница 10: ...x ...
Страница 38: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 28 Intentionally blank ...
Страница 50: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 40 Intentionally blank ...
Страница 84: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 74 Intentionally blank ...
Страница 108: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 98 Intentionally blank ...
Страница 112: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 102 Intentionally blank ...