931040P / 1014
2. THE DETECTIVE-EX/DX-100
17
Figure 12. Power Source Icons.
!
The docking station or PAC-II provides external power to simultaneously start and
operate the Detective-EX, as well as charge the internal battery. In addition, the docking
station fans help dissipate heat from the cooler. Both charger bases can be powered from
mains or a 12 V source.
!
Other external power (supplied by external battery belt or 12 V automobile battery) can
start, cool, and operate the Detective-EX, but does not charge the internal battery. The
internal battery is charged only if the external source also powers the docking station or
PAC-II. If extended remote operate is needed, use the external battery first, then
the internal battery.
The icon on the right side of the screen above the buttons
(Fig. 12) shows the current type of power connection. These
are, clockwise from upper left: (1) internal battery, (2) unit
properly connected to the docking station or power adapter/
charger, (3) external power to operate unit, and (4) battery
charging.
2.6.1. A Note on Internal vs External Battery Power
The recharging circuit for the internal battery is separate from external power so the external bat-
teries are not consumed charging the internal battery. The internal battery is only recharged
through the use of a special lithium-ion battery charger unit built into the docking station and
PAC-II using connections in the multi-pin power plug. The charging base will simultaneously
supply both “external power” and connect to the battery charger to recharge the internal battery.
When connected to an external battery, the Detective-EX relies on external power until the
external voltage drops below 9 V. It then switches to the internal battery. The instrument makes
the transition smoothly, and there is no effect on the energy resolution. If you begin with both
batteries fully charged, the sum of the two battery capacities will be achieved. If the external
battery has its own charge indicator, this can be used to determine when the changeover will
occur. The Detective-EX battery monitor indicates the available capacity in the internal battery,
and when the switch over occurs, this indicator runs down in the usual manner.
NOTE
When all power is exhausted, the high voltage and cooler will shut down and the
detector will begin to warm. The Detective-EX cannot be used again until the unit has
cooled enough that the high voltage automatically switches on and the Status
indicator says
READY
.
Содержание Detective-DX-100T
Страница 4: ...iv Intentionally blank ...
Страница 10: ...x ...
Страница 38: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 28 Intentionally blank ...
Страница 50: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 40 Intentionally blank ...
Страница 84: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 74 Intentionally blank ...
Страница 108: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 98 Intentionally blank ...
Страница 112: ...ORTEC Detective EX DX 100T Portable HPGe Spectrometer User s Manual 931040P 1014 102 Intentionally blank ...