– 60 –
APPENDIX
Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments
listed below, and should only be used in such environments:
Emissions test
Compliance
Środowisko elektromagnetyczne – wytyczne
RF emissions
CISPR 11
Group 1
RF energy is used only to maintain device’s operation. There
-
fore, its RF emissions are so low that it’s not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all establishments, including
domestic establishments, and those directly connected to the
public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments
listed below, and should only be used in such environments:
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
discharge
± 15 kV air discharge
± 8 kV contact
discharge
± 15 kV air discharge
In the case of air
discharge testing, the
climatic conditions
shall be within the
following ranges:
Ambient Temperatu
-
re:15°C~35°C, Relative
Humidity:30%~60%.
Power frequency
(50 or 60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 or 60 Hz
30 A/m
50 or 60 Hz
Power frequency ma
-
gnetic fields should be
at levels characteristic
of a typical location in
a typical commercial or
hospital environment.