– 55 –
Excessive Body Motion
Detector
Notice:
The measured blood
pressure reading may not
be accurate if the icon is
displayed.
Body movement during
measurement, especially, the
movement on the arm the
blood pressure monitor is
worn on.
e.g. Talking, moving or shaking
of the arm with the cuff on
while measurement.
Measure again. Keep arm steady
during measurement.
Cuff is worn improperly, or
the shape of the upper arm
is unusual (for example, the
circumference of the upper
arm differs largely from the
circumference of the forearm),
excessive gap might be exist
between the arm cuff and
the arm.
Wrap the cuff properly and keep
steady. Measure again.
If you have any question about the
cuff wearing and/or measurement
result, please consult your heal
-
thcare professional.
Cuff Fitness detection
Symbol
The cuff was wrapped incor
-
rectly (for example too loosely
or too tightly).
Please reference “applying the
Cuff” section to wrap the cuff
correctly.
Note: If „EP” appears on the display, just return the device to your local
distributor or importer.
LIMITED WARRANTY
Warranty For One Year from the manufacturing date
Please note that this warranty does not cover damage caused by
misuse or abuse; accident; the attachment of any unauthorized ac
-
cessory; alteration to the product; improper installation; unauthori
-
zed repairs or modifications; improper use of electrical/power sup
-
ply; loss of power; dropped product; malfunction or damage of an
operating part from failure to provide manufacturer’s recommended
maintenance; transportation damage; theft; neglect; vandalism; or
environmental conditions; loss of use during the period the product is
at a repair facility or otherwise awaiting parts or repair; or any other
conditions whatsoever that are beyond the control of importers or
distributors.
In case it is needed to have the device checked for calibration, please
consult the distributor.