3.1.1 

Input Channel 

The SR810 Lock-

In  Amplifiers  have  differential  inputs  with  6  nV/√Hz  input  noise.  The input 

impedance is 10 MΩ, and minimum full-scale input voltage sensitivity is 2 nV.  The input can also 
be configured for current measurements with selectable current gains of 106 and 108 V/A.  A line 
filter (50 Hz or 60 Hz) and a 2× line filter (100 Hz or 120 Hz) are provided to eliminate line related 
interference.  However, unlike conventional lock-in amplifiers, no tracking band-pass filter is 
needed at the input.  This filter is used by conventional lock-ins to increase dynamic reserve.  
Unfortunately, band pass filters also introduce noise, amplitude and phase error, as well as drift.  
The DSP based design of these lock-ins has such inherently large dynamic reserve that no 
tracking band-pass filter is necessary. 

 

3.1.2 

Extended Dynamic Reserve 

The dynamic reserve of a lock-in amplifier at a given full-scale input voltage is the ratio (in dB) of 
the largest interfering signal to the full-scale input voltage.  The largest interfering signal is 
defined as the amplitude  of the largest signal at any frequency  that can be applied to the input 
before the lock-in cannot measure a signal with its specified accuracy. 
 
Conventional lock-in amplifiers use an analog demodulator to mix an input signal with a reference 
signal.  Dynamic reserve is limited to about 60  dB and these instruments suffer from poor 
stability, output drift, and excessive gain  and phase  error.  Demodulation in the SR810  Lock-In 
Amplifier is accomplished by sampling the input signal with a high-precision  A/D converter, and 
multiplying the digitized input by a synthesized reference signal.  This digital demodulation 
technique results in more than 100 dB of true dynamic reserve (no prefiltering).  It is free of the 
errors associated with analog instruments. 

 

3.1.3 

Digital Filtering 

The digital signal processor also handles the task of output filtering, allowing time constants from 
10 µsec to 30,000 s, with a choice of 6, 12, 18 and 24 dB/oct roll-off.  For low frequency 
measurements (below 200 Hz), synchronous filters can be engaged to notch out multiples of the 
reference  frequency.  Since the harmonics of the reference have been eliminated, (notably 2F), 
effective output filtering can be achieved with much shorter time constants. 

 

3.1.4 

Digital Phase Shifting 

Analog phase shifting circuits have also been replaced with a DSP calculation.  Phase is 
measured with 0.01° resolution, and the X and Y outputs are orthogonal to 0.001°. 

 

3.1.5 

Frequency Synthesizer 

The built-in direct digital synthesis (DDS) source generates a very low distortion (-80 dBc) 
reference signal.  Single frequency sine waves can be generated from 1 MHz to 102 kHz with 4½ 
digits of resolution.  Both frequency and amplitude  can be set from the front panel or from a 
computer.  When using an external reference, the synthesized source is phase  locked to the 
reference signal. 

 
 

 

Содержание LIDA-SRS-KIT

Страница 1: ...Family of Brands ILX Lightwave New Focus Ophir Corion Richardson Gratings Spectra Physics MLIDA Rev A s Lock In Digital Amplifier Radiometry System User s Manual LIDA SRS KIT...

Страница 2: ...11 4 3 UNPACKING 11 4 4 CHOOSING A LOCATION 12 4 5 CHOPPER CONTROLLER CONNECTIONS 12 4 6 COMPUTER CONNECTIONS 13 4 7 DETECTOR CONNECTIONS 14 4 8 ELECTRICAL CONNECTIONS 16 5 SOFTWARE CONFIGURATION 17...

Страница 3: ...ch Locations 16 Figure 10 SR810 GPIB Address Selection 17 Figure 11 View GPIB Address on SR810 18 Figure 12 LIDA Communication Established 18 Figure 13 Setting Operating Parameters in TracQ Basic 19 F...

Страница 4: ...th each of the components 1 1 SYMBOLS AND DEFINITIONS WARNING Situation has the potential to cause bodily harm or death CAUTION Situation has the potential to cause damage to property or equipment ELE...

Страница 5: ...al will help to conserve natural resources and ensure that the materials are recycled in a manner that protects human health and the environment PROTECTIVE CONDUCTOR TERMINAL The protective conductor...

Страница 6: ...used in a manner not specified in this manual the protection provided by this equipment may be impaired To prevent damage to equipment when replacing fuses locate and correct the problem that caused t...

Страница 7: ...gnal Oriel created an application kit LIDA SRS KIT using the Stanford Research Systems SRS Lock In Digital Amplifier LIDA model SR810 The LIDA SRS KIT includes the chopper chopper control and lock in...

Страница 8: ...ference channels Reference input corrects your measurement channel to compensate for source drift TTL OUT controls the external chopper PREAMP connection may be used to power up detector pre amplifier...

Страница 9: ...with a reference signal Dynamic reserve is limited to about 60 dB and these instruments suffer from poor stability output drift and excessive gain and phase error Demodulation in the SR810 Lock In Amp...

Страница 10: ...age The SR810 has X and Y analog outputs rear panel that are updated at 256 kHz Four auxiliary inputs 16 bit ADCs are provided for general purpose use such as normalizing the input to source intensity...

Страница 11: ...ic software installation Optical detector Detector signal cable with BNC male connector Refer to the TracQ Basic user s manual for minimum system requirements for the computer If an Oriel LIDA compati...

Страница 12: ...ped light signals from an optical detector Therefore it is required to install the 75160NF Optical Chopper System Install a wheel into the chopper head based on the desired chopping frequency and beam...

Страница 13: ...nts Connect the cable to the GPIB connector on the back of the LIDA and tighten the connections Plug the USB connector into the computer If successfully installed the computer s Device Manager will di...

Страница 14: ...rs the 70310 stand alone current preamplifier to use with unamplified detectors or photomultiplier tubes Oriel s amplified detectors and stand alone current preamplifier require a power source to oper...

Страница 15: ...Richardson Gratings Spectra Physics MLIDA Rev A Figure 6 Detector Amplifier Cable Model CBL 70054 LIDA Figure 7 Amplifier Power Connection on SR810 with LIDA Compatible Oriel Detector Figure 8 Stand...

Страница 16: ...severe injury or death Refer to the specifications section of this users manual for the power requirements for the lock in digital amplifier Refer to the specifications section of the 75160NF user s m...

Страница 17: ...select the detection instrument and monochromator for the system If no monochromator instrument is to be used click Cancel Otherwise select the instrument based on the instructions provided with the...

Страница 18: ...Family of Brands ILX Lightwave New Focus Ophir Corion Richardson Gratings Spectra Physics MLIDA Rev A Figure 11 View GPIB Address on SR810 Figure 12 LIDA Communication Established...

Страница 19: ...n Richardson Gratings Spectra Physics MLIDA Rev A 5 2 SETTING OPERATING PARAMETERS To set the chopping frequency and other operating parameters for the SR810 select Detection Instr Setup Parameters Fi...

Страница 20: ...e front panel of the instrument until the INTERNAL LED is active Figure 14 Source Button SINE out and Frequency Selector Knob 6 2 SET SIGNAL INPUT MODE Set the SR810 for current mode operation by pres...

Страница 21: ...indicator OVLD next to the digital LED display Press the AUTOGAIN and AUTOPHASE buttons on the front panel of the SR810 The output LED display indicator will show the RMS current detected by the unam...

Страница 22: ...utilized for reference if its signal level is compatible with the limited dynamic range of the reference channel The range is 10 5 V to 10 5 V and the resolution is 16 bits 1 3 mV The input impedance...

Страница 23: ...Family of Brands ILX Lightwave New Focus Ophir Corion Richardson Gratings Spectra Physics MLIDA Rev A Figure 18 Reference Display Figure 19 Ratio Measurement Configuration...

Страница 24: ...the LIDA SRS KIT is compatible with a wide variety of detectors the precise procedure for obtaining calibrated results varies It is typically necessary to load wavelength and responsivity values into...

Страница 25: ...ep size interval need to be set Once this information is entered into the software the detector radiometry scan allows radiant measurement data to be taken over the detector s calibrated wavelength ra...

Страница 26: ...in real time Items needed to build the setup shown in Figure 23 LIDA SRS KIT Monochromator example Oriel model CS260 USB Q FH A Amplified calibrated detector example Oriel model DET L SIUV R C Sample...

Страница 27: ...Family of Brands ILX Lightwave New Focus Ophir Corion Richardson Gratings Spectra Physics MLIDA Rev A Figure 24 Reference Spectra Measurement Using TracQ Basic Figure 25 QE Measurements of Samples...

Страница 28: ...n DET L SIUV R C UV enhanced Si photodiode w transimpedance amplifier 200 1100 nm sensitivity 10 mm x 10 mm active area NIST traceable calibration DET L SIUV T C UV enhanced Si photodiode w integrated...

Страница 29: ...readings Many of Oriel s legacy amplified detectors may easily be integrated with the LIDA provided the detector has Mini DIN and BNC connectors Discard the Mini DIN connector included with the legac...

Страница 30: ...y 5 ppm C REFERENCE CHANNEL Frequency Range 0 001 Hz to 102 4 kHz hopper wheel dependent Reference Input TTL or sine 400 mVpp min Input Impedance 1 M 25 pF Phase Resolution 0 01 front panel 0 008 thro...

Страница 31: ...1 Output 10 V output of X R X noise Aux 1 or Aux 2 Updated at 512 Hz X Y Outputs In phase and quadrature components rear panel 10 V updated at 256 kHz Aux A D Inputs 4 BNC inputs 10 V 1 mV resolution...

Страница 32: ...l subsidiaries and sales offices worldwide Every year the Newport Resource catalog is hailed as the premier sourcebook for those in need of advanced technology products and services It is available by...

Страница 33: ...his RMA number must appear on both the shipping container and the package documents Return the product to Newport freight prepaid clearly marked with the RMA number and it either will be repaired or r...

Страница 34: ...Newport in writing promptly after discovery of the defect or nonconformity and within the Warranty Period Products may only be returned by Buyer when accompanied by a return material authorization num...

Страница 35: ...Materials contain trade secrets of Newport and are for use only in the manner expressly permitted Newport claims and reserves all rights and benefits afforded under law in the Programs provided by New...

Отзывы: