Family of Brands – ILX Lightwave
®
• New Focus™ • Ophir
®
• Corion • Richardson Gratings™ • Spectra-Physics
®
MLIDA, Rev A
Range
1 MHz to 102 kHz
Accuracy
25 ppm + 30 µHz
Frequency Resolution
4-1/2 digits or 0.1 MHz, whichever is greater
Distortion
-80 dBc (f<10 kHz), -70 dBc (f> 10 kHz) @ 1 Vrms
amplitude
Amplitude
0.004 to 5 Vrms into 10
kΩ (2
mV resolution), 50
Ω
output impedance, 50 mA maximum current into 50
Ω
Amplitude Accuracy
1%
Amplitude Stability
50 ppm/°C
Outputs
Sine, TTL
When using an external reference, both outputs are
phase locked to the external reference.
DISPLAYS
Channel 1 Output
±10 V output of X, R, X-noise, Aux 1 or Aux 2.
Updated at 512 Hz.
Offset
X, Y, R can be offset up to ±105% of full scale
Expand
X, Y, R can be expanded by 10x or 100x
Reference
4-1/2 digit LED display
INPUTS AND OUTPUTS
Channel 1 Output
±10 V output of X, R, X-noise, Aux 1 or Aux 2.
Updated at 512 Hz.
X, Y Outputs
In-phase and quadrature components
(rear panel)
(±10 V), updated at 256 kHz
Aux. A/D Inputs
4 BNC inputs, ±10 V, 1 mV resolution, sampled at
512 Hz
Aux. D/A Outputs
4 BNC outputs, ±10 V, 1 mV resolution
Sine Out
Internal oscillator analog output
TTL Out
Internal oscillator TTL output
Data Buffer
8k point buffer with data recorded at rates to 512 Hz
and read through the computer interfaces.
Trigger In (TTL)
Trigger synchronizes data recording
Remote Pre-amp
Provides power to the preamplifier integrated inside
Oriel's LIDA detectors, using cable Model CBL-
70054-LIDA
Computer Interface
IEEE-488.2 and RS-232 interfaces standard. All
instrument functions can be controlled and read
through IEEE-488.2 or RS-232 interfaces. TracQ Basic
software utilizes GPIB communication.
GENERAL
Power
100 to 120 VAC, 220 to 240 VAC, 50 to 60 Hz, 40 W
Dimensions
17" W x 5.25" H x 19.5" L [431 x 133 x 495 mm]
Weight
30 lbs [13.6 kg]