DP-100SL
35
Site Preparation Guide
Chapter 8
Air
Compressed Air
shows the requirements for compressed air at the site.
It is the customer’s responsibility to maintain filtering of the outside
supply to ensure an oil-free compressed air supply.
Table 8.1 Compressed Air Requirements
Pressure
0.7 ~ 1.0 MPa (7 - 10 bar)
Air supply rate
55 liters/minute
Quality
Dry (water: 0.03 g/m³)
Filtered (dust: 0.1 µm, 0.1 mg/m³)
Free of oil (oil: 0.01 mg/m³)
Tube connectors
10 mm (outside diameter)
Содержание DP-100SL
Страница 1: ...DP 100SL Site Preparation Guide October 2003 Revision A ...
Страница 2: ......
Страница 14: ......
Страница 18: ......
Страница 25: ...DP 100SL 19 Site Preparation Guide Figure 3 2 System Dimensions Front View ...
Страница 26: ...20 Chapter 3 Facility Physical Requirements Figure 3 3 System Dimensions Side View ...
Страница 27: ...DP 100SL 21 Site Preparation Guide Figure 3 4 Site Dimensions Standard Configuration ...
Страница 32: ......
Страница 40: ......
Страница 48: ......