DP-100SL
25
Site Preparation Guide
Humidity
* Storage humidity allowed only when in original packaging.
Heat Dissipation
Ambient Air Conditions
You should install:
• A fresh air outlet above the laser supply box (i.e. power heat
sink).
• A warm air exhaust close to the main heat source (i.e. laser supply box).
Take note that:
• No organic chemicals allowed in the air.
• No ammonia allowed in the air.
Table 4.2 Operating Humidity
Unit
Operating Humidity
Storage Humidity*
DP-100SL
50% to 60%
10% to 80%, non-condensing
Table 4.3 Maximum Heat Dissipation for System Units
Unit
Heat Dissipation
DP-100SL
5 kW (228 BTU/minute) - To air
Pump
2.5 kW (114 BTU/minute) - To air
Содержание DP-100SL