78
-
5.2 Autonomous operation Xi 80/410
A special feature of the Xi 80/410 is the autonomous operation. There is no need for a permanent connection
to the PIX Connect software. Only a few settings must be set in advance in the software.
To do this, connect the PIF and Ethernet or USB cables
to the device. Then connect the Xi to your PC and start
the PIX Connect software (see
) the
camera so that your object to be measured is perfectly
visible in the image. First define the desired
measurement area with the corresponding mode you
want to output.
Figure 36:
Connection Xi 80/410 to PC via
Ethernet/USB
Now go to the configuration menu on
Device (PIF)
. There you first select the
PIF type
(in this case:
Internal
PIF
). Then select under
Analog Outputs (AO)
the function that is to be output autonomously. Then press
Setup
, make your settings, and make sure that the checkbox is set to
using autonomously by device
. By
pressing the
OK
button an
@
sign can be found in the configuration menu Devices (PIF), which indicates the
autonomous operation.
Содержание Xi 400
Страница 1: ...Operator s Manual optris Xi 80 400 410 Spot finder IR camera...
Страница 27: ...Mechanical Installation 27 Figure 5 Xi 80 dimensions mm...
Страница 28: ...28 Figure 6 Xi 400 dimensions mm...
Страница 29: ...Mechanical Installation 29 Figure 7 Xi 410 dimensions mm...
Страница 30: ...30 Figure 8 Industrial stackable PIF Process Interface electronic box control box shutter dimensions mm...
Страница 34: ...34 Figure 11 Mounting bracket ACXIAPLAB dimensions mm weight 276 g...
Страница 36: ...36 Figure 12 Water cooling ACXIW and mounting kit ACXIxxxWAK1 dimensions mm weight 1710 g...
Страница 37: ...Mechanical Installation 37 Figure 13 Shutter for water cooling ACXISW dimensions mm weight 600 g...
Страница 39: ...Mechanical Installation 39 Figure 14 Shutter ACXIS with mounting bracket ACXIAPLAB dimensions mm weight 826 g...
Страница 40: ...40 Figure 15 Shutter ACXIS dimensions mm weight 550 g...
Страница 71: ...Functions 71 Figure 35 Possibilities of power supply for Xi 80 410 via Ethernet connection...
Страница 79: ...Functions 79 Figure 37 Configurations menu Device PIF Figure 38 Setup...
Страница 92: ...92 7 2 Software window Figure 45 Software window 1 2 3 2 5 6 7 9 4 10 11 12 1 8...
Страница 107: ...Appendix A Table of emissivity for metals 107 Appendix A Table of emissivity for metals...
Страница 108: ...108...
Страница 109: ...Appendix B Table of emissivity for non metals 109 Appendix B Table of emissivity for non metals...
Страница 118: ...118 Figure 58 Relay output at industrial PIF...
Страница 119: ...Appendix G Declaration of Conformity 119 Appendix G Declaration of Conformity...
Страница 120: ...optris Xi MA E2021 08 A...