RS485 for Xi 80 and Xi 410
The Xi 80 and Xi 410 cameras support a direct RS485 connection. RS485 interface is designed for high-
speed serial data transmission over long distances and as a bidirectional bus system with up to 32
participants. An RS485 bus can be set up as a 2-wire as well as a 4-wire system. For the Xi 80/410 cameras,
it is designed for a 2-wire system. A maximum cable length of 500 meters can be realized. At least a master
and slave is necessary for communication. The master can be a PLC system and the slave is the camera.
RS485 is supported for Xi 80 cameras which have the hardware version 3001, firmware 3023 and at least
the serial number which starts with 1812xxxx. For the Xi 410 it is working for all cameras starting at firmware
version 3815. Here it is important that the communication is just working via Ethernet connection or in
autonomous mode, but not with USB.
The connection for RS485 can be done directly on the terminal block (A and B) which is provided in the
scope of supply of the camera.
Command list
The command list and more information can be found on the supplied USB stick under
Documentation and Manuals (Serial Communication Description-RS485-xxxx-xx-x). Every
command must expire with CR/LF (0x0D, 0x0A).
Содержание Xi 400
Страница 1: ...Operator s Manual optris Xi 80 400 410 Spot finder IR camera...
Страница 27: ...Mechanical Installation 27 Figure 5 Xi 80 dimensions mm...
Страница 28: ...28 Figure 6 Xi 400 dimensions mm...
Страница 29: ...Mechanical Installation 29 Figure 7 Xi 410 dimensions mm...
Страница 30: ...30 Figure 8 Industrial stackable PIF Process Interface electronic box control box shutter dimensions mm...
Страница 34: ...34 Figure 11 Mounting bracket ACXIAPLAB dimensions mm weight 276 g...
Страница 36: ...36 Figure 12 Water cooling ACXIW and mounting kit ACXIxxxWAK1 dimensions mm weight 1710 g...
Страница 37: ...Mechanical Installation 37 Figure 13 Shutter for water cooling ACXISW dimensions mm weight 600 g...
Страница 39: ...Mechanical Installation 39 Figure 14 Shutter ACXIS with mounting bracket ACXIAPLAB dimensions mm weight 826 g...
Страница 40: ...40 Figure 15 Shutter ACXIS dimensions mm weight 550 g...
Страница 71: ...Functions 71 Figure 35 Possibilities of power supply for Xi 80 410 via Ethernet connection...
Страница 79: ...Functions 79 Figure 37 Configurations menu Device PIF Figure 38 Setup...
Страница 92: ...92 7 2 Software window Figure 45 Software window 1 2 3 2 5 6 7 9 4 10 11 12 1 8...
Страница 107: ...Appendix A Table of emissivity for metals 107 Appendix A Table of emissivity for metals...
Страница 108: ...108...
Страница 109: ...Appendix B Table of emissivity for non metals 109 Appendix B Table of emissivity for non metals...
Страница 118: ...118 Figure 58 Relay output at industrial PIF...
Страница 119: ...Appendix G Declaration of Conformity 119 Appendix G Declaration of Conformity...
Страница 120: ...optris Xi MA E2021 08 A...