20
-
■
HFOV
: Horizontal enlargement of the total measuring at object level
■
VFOV
: Vertical enlargement of the total measuring at object level
■
IFOV
: Size at the single pixel at object level
■
DFOV
: Diagonal dimension of the total measuring field at object level
■
MFOV
: Recommended, smallest measured object size of 3 x 3 pixel
(Xi 400/410) and 2 x 2 pixel (Xi 80)
Geometric resolution for ideal temperature measurement
When designing optics for measuring IR cameras, special attention must be paid to the quality of detail
contrast with which an object can be represented in the image. This is described by the modulation transfer
function (MTF). Since in contrast to visual cameras, with IR cameras the thermal contrast is of more interest,
this is used together with the slit response function (SRF). The result is determined by the number of pixels
an object needs to fill to allow its temperature to be measured exactly. In high-performance infrared optical
systems as used by Optris, this is
3×3 pixels
or
2x2 pixels
, with lower quality optical systems, in some
circumstances as many as 10×10 pixels may be required, to receive
90 %
of the energy. A high-performance
Содержание Xi 400
Страница 1: ...Operator s Manual optris Xi 80 400 410 Spot finder IR camera...
Страница 27: ...Mechanical Installation 27 Figure 5 Xi 80 dimensions mm...
Страница 28: ...28 Figure 6 Xi 400 dimensions mm...
Страница 29: ...Mechanical Installation 29 Figure 7 Xi 410 dimensions mm...
Страница 30: ...30 Figure 8 Industrial stackable PIF Process Interface electronic box control box shutter dimensions mm...
Страница 34: ...34 Figure 11 Mounting bracket ACXIAPLAB dimensions mm weight 276 g...
Страница 36: ...36 Figure 12 Water cooling ACXIW and mounting kit ACXIxxxWAK1 dimensions mm weight 1710 g...
Страница 37: ...Mechanical Installation 37 Figure 13 Shutter for water cooling ACXISW dimensions mm weight 600 g...
Страница 39: ...Mechanical Installation 39 Figure 14 Shutter ACXIS with mounting bracket ACXIAPLAB dimensions mm weight 826 g...
Страница 40: ...40 Figure 15 Shutter ACXIS dimensions mm weight 550 g...
Страница 71: ...Functions 71 Figure 35 Possibilities of power supply for Xi 80 410 via Ethernet connection...
Страница 79: ...Functions 79 Figure 37 Configurations menu Device PIF Figure 38 Setup...
Страница 92: ...92 7 2 Software window Figure 45 Software window 1 2 3 2 5 6 7 9 4 10 11 12 1 8...
Страница 107: ...Appendix A Table of emissivity for metals 107 Appendix A Table of emissivity for metals...
Страница 108: ...108...
Страница 109: ...Appendix B Table of emissivity for non metals 109 Appendix B Table of emissivity for non metals...
Страница 118: ...118 Figure 58 Relay output at industrial PIF...
Страница 119: ...Appendix G Declaration of Conformity 119 Appendix G Declaration of Conformity...
Страница 120: ...optris Xi MA E2021 08 A...