package. Pay attention to moisture and vibration. If the instrument is to be stored for a long period, it 

shall be cleaned and put into the package and kept in a dry and well ventilated store room without 

any  corrosive  chemicals.  The  relative  humidity  of  the  room  shall  not  exceed  80%  and  the 

temperature shall be controlled in the range of -40 ° C - 55 ° C without any temperature shock.   

 

10. Trouble-shooting Guide   

In case there is any trouble, please first refer to the trouble-shooting guide. If it still cannot work, 

please contact the authorized distributor.   

 

Trouble 

Possible reason 

Remedy 

No working 

Power supply adapter is not connected 

Connect the power supply adapter 

No switch on the main power 

Switch on the main power 

Main power is broken-down 

Contact a local electrician 

Illumination 

does not light 

The LED lamp is not connected correctly 

Connect again 

No switch on the main power 

Switch on the main power 

Main power is broken-down 

Contact a local electrician 

Electric parts are broken-down 

Contact the authorized distributor 

The bulb is burnt 

Replace the spare bulb 

 

11. Fault Checking and Solution for LED illumination System.   

 

After installation, power-

on test according to user’s manual, or during using operation microscope, find 

no light, please check according to the following steps: 

 

11.1 Connect power supply adapter to power supply and power supply box. By this time, the indicator 

light of power supply adapter is always on. Turn on the power switch on the power box. The power input 

is normal if the switch indicator is on. The power input is abnormal if the switch indicator is off, please 

check:  A.  Connection  of  the  power  supply  adapter  again.  B.  Where  the  power  supply  adapter  is 

damaged, if yes, please replace. 

 

 

11.2  When  the  switch  indicator  is  on,  turn  brightness  control  knob  to  the  maximum,  and  checks  the 

Содержание A41.1903

Страница 1: ...safety and obtain satisfactory performance please study this instruction manual thoroughly before start to use your instrument A41 1903 Operating Microscope One Head 0 180 Manual Step Zoom 4 8x 16x F...

Страница 2: ...agnification change Fine focusing can be done by objective fine adjusting lever Two objectives are available for different working distance and magnification The illumination system of the instrument...

Страница 3: ...ng to avoid accident When moving the instrument in the room it is recommended to push it at the lower part to prevent the instrument from being upset and damaged 1 4 Do not touch the surface of the le...

Страница 4: ...5 Diameter of Spot 57 00 mm 71 00 mm 86 00 mm 113 00 mm Exit Pupil Distance 25 00 mm 2 4 Table of Specification for A41 1903 Operation Microscope Model A41 1903 Binocular Head Tilting 0 180 Main Usag...

Страница 5: ...6 Arm assembly 7 0 180 Binocular 8 Eyepiece 9 CCD camera optional 10 Demonstrator optional 11 Beamsplitter optional 12 Magnification changer 13 Video camera adapter optional 4 Working Principle and Fu...

Страница 6: ...socket for camera 6 Brightness control knob 7 Power switch Note Non professional maintenance staffs do not open the power supply box 4 3 Arm assembly see Figure 1 The arm assembly is an important part...

Страница 7: ...r which connects with beamsplitter 3 Focusing hand wheel 4 Locking screw8 c Demonstrator 10 optional It is a monocular observation unit connected to the beamsplitter It can be used for observation at...

Страница 8: ...objective you need Turn fine focusing lever counterclockwise to make the fine focusing lever in the middle position 5 7 Installation of operation handle Fix the operation handle on the head of micros...

Страница 9: ...peat step 7 3 7 5 Step down the foot wheel brake during operation to present the instrument from being moved 8 Care and Maintenance The instrument shall be placed in a dry dust free and well ventilate...

Страница 10: ...on does not light The LED lamp is not connected correctly Connect again No switch on the main power Switch on the main power Main power is broken down Contact a local electrician Electric parts are br...

Страница 11: ...t LED lamp plug into the DC output socket 11 3 Remove four screws in red circle open under cover from power supply box Press the plug connectors in LED drive circuit board and repeat step 2 If not DC...

Страница 12: ...connection of power line from power supply box to LED bulb If yes please replace LED bulb 11 6 If you have two operation microscopes in your hand you can cross check for example install this microscop...

Страница 13: ...ose wire clip C Remove two screws fixed LED bulb and separate LED bulb from LED bulb seat Welding off two electric wire from LED bulb and remember the welding position for each electronic wire Weld tw...