delivered. If the used finds the torque not suitable (too loose or too tight when turning), it can be 

adjusted by special person in the following way: 

 

6.2.1 Screw off locking knob 1, screw 2 and nut 3. Tighten nut 4 to increase torque or vice versa. 

Then screw down nut 3, tighten screw 2 and screw 2 and screw in knob 1.   

 

6.2.2 The torque adjustment of locking knob 6 is the same as that of locking knob 1. The torque is 

adjusted by nut 7.   

 

7. Operation of instrument 

 

7.1  Connect  the  power  supply  adapter  with  power  supply  box,  and  connect  the  power  supply 

adapter with power socket 

 

7.2  Press  power  switch.  Turn  dimmer  knob  for  proper  brightness.  Note:  Before  switching  off  the 

power,  make  sure  that  the  brightness  is  adjusted  to  minimum  to  prevent  the  lamp  from  being 

damaged when switching on again.   

 

7.3 Hold handle, align the microscope with the object to be observed and move the microscope up 

and down, left and right till a distinct image is seen by the operator (The eyecup can be rolled down 

if the observer wears glasses). Then tighten handwheel. Fine focusing can be done by fine focusing 

lever during operation.   

 

7.4  If  the  objective  is  to  be  replaced  for  magnification  change,  refocusing  will  be  needed  after 

replacement. (i.e. repeat step 7.3)   

7.5 Step down the foot wheel brake during operation to present the instrument from being moved.   

 

8. Care and Maintenance   

 

The  instrument  shall  be  placed  in  a  dry,  dust-free  and  well-ventilated  room  where  no  corrosive 

chemicals  are  stored.  The  instrument  shall  not  be  disassembled  by  non-professional  people, 

especially the optical part. Never touch the glass surface with finger. If the dust, dirt and oil on the 

glass surface is to be cleaned, first clean the dust off the surface with a clean and soft brush, then 

wipe  the  surface  gently  with  degreased  cotton  soaked  with  a  little  absolute  alcohol.  The  glass 

surface  shall  not  be  wiped  frequently  to  avoid  damage  to  the  surface.  The  instrument  shall  be 

covered with clean cloth when it is not in use.   

 

Often by the relative moisture shall be less than 80%, temperature shall be within 5 - 40ºC without 

temperature  shock.  It  is  forbidden  to  disassemble  the  instrument  by  an  unqualified  person, 

especially for optical parts, Do not touch the glass surface with finger directly. When cleaning, wipe 

the  glass  surface  with  absorbent  cotton  slightly  moistened  with  absolute  alcohol  after  the  glass 

surface has been cleaned of dust with a clean and soft brush. Wipe it as few times as possible to 

avoid damage to the glass surface. Cover the instrument with a clean cloth cap when it lies in idle.   

 

9. Transportation and Storage 

During  transportation,  the  instrument  shall  be  placed  according  to  the  indication  marked  on  the 

Содержание A41.1903

Страница 1: ...safety and obtain satisfactory performance please study this instruction manual thoroughly before start to use your instrument A41 1903 Operating Microscope One Head 0 180 Manual Step Zoom 4 8x 16x F...

Страница 2: ...agnification change Fine focusing can be done by objective fine adjusting lever Two objectives are available for different working distance and magnification The illumination system of the instrument...

Страница 3: ...ng to avoid accident When moving the instrument in the room it is recommended to push it at the lower part to prevent the instrument from being upset and damaged 1 4 Do not touch the surface of the le...

Страница 4: ...5 Diameter of Spot 57 00 mm 71 00 mm 86 00 mm 113 00 mm Exit Pupil Distance 25 00 mm 2 4 Table of Specification for A41 1903 Operation Microscope Model A41 1903 Binocular Head Tilting 0 180 Main Usag...

Страница 5: ...6 Arm assembly 7 0 180 Binocular 8 Eyepiece 9 CCD camera optional 10 Demonstrator optional 11 Beamsplitter optional 12 Magnification changer 13 Video camera adapter optional 4 Working Principle and Fu...

Страница 6: ...socket for camera 6 Brightness control knob 7 Power switch Note Non professional maintenance staffs do not open the power supply box 4 3 Arm assembly see Figure 1 The arm assembly is an important part...

Страница 7: ...r which connects with beamsplitter 3 Focusing hand wheel 4 Locking screw8 c Demonstrator 10 optional It is a monocular observation unit connected to the beamsplitter It can be used for observation at...

Страница 8: ...objective you need Turn fine focusing lever counterclockwise to make the fine focusing lever in the middle position 5 7 Installation of operation handle Fix the operation handle on the head of micros...

Страница 9: ...peat step 7 3 7 5 Step down the foot wheel brake during operation to present the instrument from being moved 8 Care and Maintenance The instrument shall be placed in a dry dust free and well ventilate...

Страница 10: ...on does not light The LED lamp is not connected correctly Connect again No switch on the main power Switch on the main power Main power is broken down Contact a local electrician Electric parts are br...

Страница 11: ...t LED lamp plug into the DC output socket 11 3 Remove four screws in red circle open under cover from power supply box Press the plug connectors in LED drive circuit board and repeat step 2 If not DC...

Страница 12: ...connection of power line from power supply box to LED bulb If yes please replace LED bulb 11 6 If you have two operation microscopes in your hand you can cross check for example install this microscop...

Страница 13: ...ose wire clip C Remove two screws fixed LED bulb and separate LED bulb from LED bulb seat Welding off two electric wire from LED bulb and remember the welding position for each electronic wire Weld tw...

Страница 14: ...ter optional 1 pc 4 CCD camera optional 1 pc 5 Demonstrator optional 1 pc 6 Objectives F 200mm fitted on the microscope 1 pc F 250mm 1 pc F 300mm optional 1 pc F 400mm optional 1 pc 7 Power supply box...

Отзывы: