Chapter 6 Objectives
A16.1098
c
、
Ensure that the matched phase
contrast ring 2(in the phase
contrast objective) and phase
contrast ring 1(in condenser) have
been moved into optical path.
d
、
Observing by centering
telescope,adjust two centering
adjustment screws in figure 26
with 1.5mm hexagonal screw until
the
centers of two phase contrast rings
overlap.
e
、
The ways above are suitable for
phase contrast objectives with
various ratio.
★
If the phase contrast rings are
not accurately centered,the best
effect
of
phase
contrast
observation will not be obtained.
★
Phase contrast ring may have
deviation after moving or changing
a thicker sample.If so,the steps
above are needed to adjust until the
centers of two phase contrast rings
overlap.
★
If the glass slide or vessel is not
flat,to achieve greater contrast,the
steps above are also needed to
adjust,until the centers of two
phase contrast rings overlap.