18
Chapter 3.Installation
A16.1098
3.10 Multi-function modes
(
Figure 14
&
Figure 15
)
①
Turn up the scale cover plate,
move the locking handle downward,
and pull out or insert the turntable
from the right.
②
When inserting the module, both
sides of the module shall be inserted
into the turntable along the lower
spring plate, the end face shall be
close, and the upper spring plate
shall firmly press the top of the
module
★
Note
that
the
module
identification corresponds to the
digital
identification
of
the
turntable.
3.11 Power connection
(
Figure 16
)
①
Plug the LED light source cable
into the corresponding socket
②
Wire the z-axis control port on
the control box to the port on the
rack
③
Insert the rocker connecting cable
into the corresponding port of the
control box
Figure 14
Figure 15
pocket for analyzer
cover
locking
reflector
engraved number
bottom spring clamp
reflector module
top spring clamp
Figure 16