    Resolution = λ / ( 2 

×

 N.A.) 

    λ stands for wavelength ( λ = 0.55μmm) 
 

Eyepiece view field 

      The max diameter measured when the diaphragm is open. 10X/20 means the magnification is 10X and 

the max diameter of view field is 20mm 

 

Effective view field 

      Liner view field which is observed on the object surface 
      Effective view field = eyepiece view field / objective’s magnification 
 

Depth of field 

      Depth of field stands for depth of the space in which there is sharp image on the object surface. Depth 

of  field  will  be  longer  when  the  diaphragm  reduces.  Depth  of  field  will  be  shorter  when  the  N.A. 
becomes bigger. 

3. Configuration 

Item 

Content 

Piece 

Microscope’s body ( including stage, nosepiece, 

coaxial coarse and fine focusing adjustment, 

bracket for condenser, adjusting power supply, 

halogen bulb and normal collector) 

1pc 

Siedentopf binocular inclined at 30

º

(360

º

 

rotatable) 

1pc 

Condenser ( with iris diaphragm, without 

auxiliary lens) 

1pc 

 

Objectives 

( Infinity achromatic 

objectives) 

4X 

 

1 set 

10X 
40X 

100X 

WF10X eyepiece 

2 pcs 

Blue filter 

1 set 

Electronic cord 

1 pc 

Immersion oil 

1 bottle 

Dust cover 

1 pc 

10 

Instruction and quality certificate 

1 pc 

11 

Halogen bulb ( 12V20W) 

2 pcs 

12 

Wrench 2.5, 1.5 

1pc 

 
Optional 

Item 

Contents 

Piece 

Collector with diaphragm 

 

Condenser ( with iris diaphragm and auxiliary 

lens) 

 

Dark field condenser 

 

WF10X eyepiece with reticle 

 

WF15X eyepiece (15mm) 

 

 

Objectives 

(Infinity plan 

4X 

 

10X 

 

Содержание A12.1010

Страница 1: ...A12 1010 Laboratory Biological Microscope Instruction Manual Please read this manual carefully before use microscope...

Страница 2: ...Contents BM2000 1 Structure and nomenclature 4 2 Technical parameter and specification 6 3 Configuration 8 4 Assemblage 9 5 Operation 13 6 Troubleshooting 18 7 Maintenance 20 8 Products Standards 20...

Страница 3: ...oscope on a stable plane and keep it in balance Keep the environment breezy and cover the microscope with the dust cover 2 6 Do not place microscope in moist room to avoid short circuit Please turn of...

Страница 4: ...inserted into the holder firmly Such Operation may cause short circuit or smoking At last please make sure if the collector has been mounted properly 9 Temperature for illumination There will be high...

Страница 5: ...re screw Nosepiece Objectives Specimen holder Mechanical stage Condenser with auxiliary lens Collector Binocular head Eye tubes fixture Microscope body Power supply Lightness adjusting knob switch scr...

Страница 6: ...Working environment Temperature 0 40 Humidity max 85 no dew Altitude within 2000m Pollution 2 Assembly 2 Optical parameter 2 1 Objectives infinity achromatic objectives Magnification N A Cover glass t...

Страница 7: ...tem Content Piece 1 Microscope s body including stage nosepiece coaxial coarse and fine focusing adjustment bracket for condenser adjusting power supply halogen bulb and normal collector 1pc 2 Siedent...

Страница 8: ...ded AC connector and make sure that the electronic cord has been connected safely Please use the supplied electronic cord by manufacturer Please make sure to use PE electronic cord when using extra lo...

Страница 9: ...rsion image disappeared 4 5 2 Objectives The objectives have been mounted in the microscope before shipment The objectives have been mounted in the microscope before shipment Take off the specimen fro...

Страница 10: ...lacement for fuse 4 6 2 1 Turn off the power supply turn the switch to O and pull out the connector 4 6 2 2 Open the cover for fuse by screwdriver 4 6 2 3 Mount the new fuse 4 6 2 4 Make sure that the...

Страница 11: ...there is sharp image 5 4 4 Turn the fine focusing knob to focus finely Use the 10X objective or 4X objective firstly before 40X and 100X objective Turn the fine focusing knob to focus finely 5 5 Adjus...

Страница 12: ...ctive when the aperture diaphragm adjusting shaft reaches some position If the objective has been changed please move the shaft to the position with the same value which is in the objective in the opt...

Страница 13: ...4 2 Never press the bottle overly as it may caused oil gushed out of the bottle 5 9 4 3 Clean the out surface of the bottle from remanent oil 5 10 Adjusting the coarse focusing knob s tension The tens...

Страница 14: ...or dust on the slide Cleaning Condenser position too low Correct position Poor image quality low resolution poor contrast No cover glass on the specimen Attach cover glass Cover glass too thick or thi...

Страница 15: ...Inadequate brightness or illumination Adjust brightness with control dial 3 Electrical Trouble Cause Remedy Lamp dose not light when switched ON No electrical power Check power cord connection Lamp b...

Страница 16: ...er thinner ect to clean the painted or plastic surfaces of the instrument We recommend the use of silicon cloth More persistent dirt may be cleaned with mild detergent solution Printed plastic surface...