                                As  the  air  bubble  in  the  oil  immersion  will  make  bad  affect  to  the  quality  of  image, 

please  make  sure  that  there  is  no  air  bubble  in  the  oil  immersion  before  use.  Please 
check  the  air  bubble  as  following,  remove  the  eyepieces  and  open  the  view  field 
diaphragm  and  aperture  diaphragm entirely  then observe the exit  pupil which  is light 
and in nummular shape 

                  5.9.2.3 Remove the air bubble as following 
                                Turn the nosepiece slowly and turn the oil immersion objective once or twice 
                                Turn the lifting knob for condenser slowly and make the condenser a little bit of up and 

down 

                                Increase the amount of oil immersion or change the old oil immersion with new one 
        5.9.3 Operation for oil immersion 
                    Use the oil immersion as little as possible. The oil immersion will be conglutinated to the stage 

and condenser and this will have bad effect the performance if there is too much oil immersion. 
Please clean the redundant oil on the objective and condenser after finishing the observation 
otherwise the image will be affected. Use the aether to clean the oil then use the pure alcohol 
(ethanol or carbinol) to make entirely cleaning. Please repeat the clean three or four times. 

                    Notice:  Please  follow  the  instruction  from  manufactures  when  the  aether  or  pure  alcohol  is 

used. Keep them away from the fire or electronic spark 

        5.9.4 Attention 
                  5.9.4.1 Keep the bottle for oil immersion sealed as much as possible and check it periodically 
                  5.9.4.2 Never press the bottle overly as it may caused oil gushed out of the bottle 
                  5.9.4.3 Clean the out surface of the bottle from remanent oil 
5.10 Adjusting the coarse focusing knob’s tension 
        The tension of coarse focusing knob is adjustable. Please turn the   
tension controlling ring of the coarse focusing knob clockwise then the   
tension will be increased. This ring is near the coarse focusing knob.   
If you want decrease the tension, please turn the ring withershins. The   
stage will fall down automatically if the tension of the ring decreased   
too much 
5.11 Micrography   
        The image from microscope could be get by camera, digital camera and CMOS/CCD camera 

5.11.1 Installation and operation for CMOS/CCD camera 
          5.11.1.1 Installation 

It is similar to assemble the eyepiece tube, just tighten the screw to fix it. Connect the 

CMOS or CCD camera with the adaptor and insert it in the photo tube. Adjust the image 
to be correct then tighten the screw. 

Use the USB cable to connect the CMOS or CCD camera and the computer

(Please 

read the instruction in the software to see how to use the software) 

                  5.11.1.2 Operation 
                           

Observe  through  the  10X  eyepieces  before  photography.  Do  focusing  as  the  above 

steps  till  sharp  image  of  the  specimen  displayed  by  the  software  and  do  not  do  any 
adjustment to the camera system 

                           

If  the  image  from  the  software  is  too  bright  or  dark,  please  turn  the  power  supply’s 

brightness adjusting knob to adjust brightness of the illumination 

                           

Turn the fine focusing knob to make the microscope is exactly on the focusing plane. 

Then turn the diopter adjusting ring on the eyepieces to make sure them are on focus 

Содержание A12.1010

Страница 1: ...A12 1010 Laboratory Biological Microscope Instruction Manual Please read this manual carefully before use microscope...

Страница 2: ...Contents BM2000 1 Structure and nomenclature 4 2 Technical parameter and specification 6 3 Configuration 8 4 Assemblage 9 5 Operation 13 6 Troubleshooting 18 7 Maintenance 20 8 Products Standards 20...

Страница 3: ...oscope on a stable plane and keep it in balance Keep the environment breezy and cover the microscope with the dust cover 2 6 Do not place microscope in moist room to avoid short circuit Please turn of...

Страница 4: ...inserted into the holder firmly Such Operation may cause short circuit or smoking At last please make sure if the collector has been mounted properly 9 Temperature for illumination There will be high...

Страница 5: ...re screw Nosepiece Objectives Specimen holder Mechanical stage Condenser with auxiliary lens Collector Binocular head Eye tubes fixture Microscope body Power supply Lightness adjusting knob switch scr...

Страница 6: ...Working environment Temperature 0 40 Humidity max 85 no dew Altitude within 2000m Pollution 2 Assembly 2 Optical parameter 2 1 Objectives infinity achromatic objectives Magnification N A Cover glass t...

Страница 7: ...tem Content Piece 1 Microscope s body including stage nosepiece coaxial coarse and fine focusing adjustment bracket for condenser adjusting power supply halogen bulb and normal collector 1pc 2 Siedent...

Страница 8: ...ded AC connector and make sure that the electronic cord has been connected safely Please use the supplied electronic cord by manufacturer Please make sure to use PE electronic cord when using extra lo...

Страница 9: ...rsion image disappeared 4 5 2 Objectives The objectives have been mounted in the microscope before shipment The objectives have been mounted in the microscope before shipment Take off the specimen fro...

Страница 10: ...lacement for fuse 4 6 2 1 Turn off the power supply turn the switch to O and pull out the connector 4 6 2 2 Open the cover for fuse by screwdriver 4 6 2 3 Mount the new fuse 4 6 2 4 Make sure that the...

Страница 11: ...there is sharp image 5 4 4 Turn the fine focusing knob to focus finely Use the 10X objective or 4X objective firstly before 40X and 100X objective Turn the fine focusing knob to focus finely 5 5 Adjus...

Страница 12: ...ctive when the aperture diaphragm adjusting shaft reaches some position If the objective has been changed please move the shaft to the position with the same value which is in the objective in the opt...

Страница 13: ...4 2 Never press the bottle overly as it may caused oil gushed out of the bottle 5 9 4 3 Clean the out surface of the bottle from remanent oil 5 10 Adjusting the coarse focusing knob s tension The tens...

Страница 14: ...or dust on the slide Cleaning Condenser position too low Correct position Poor image quality low resolution poor contrast No cover glass on the specimen Attach cover glass Cover glass too thick or thi...

Страница 15: ...Inadequate brightness or illumination Adjust brightness with control dial 3 Electrical Trouble Cause Remedy Lamp dose not light when switched ON No electrical power Check power cord connection Lamp b...