background image

accordingly 

    5.8 Adjusting aperture diaphragm 
          Adjust  the  aperture  diaphragm  adjusting  shaft  for  condenser  till  the  position  which  stands  for  the 

magnification of objective 

         

adjust the aperture diaphragm 

          5.8.1  Adjust  the  aperture  diaphragm  by  the  adjusting  shaft.  The  brightness  and  resolution  will  be 

decrease  while  the  contrast  and  depth  of  view  will  increase  if  the  diameter  of  aperture 
diaphragm reduces 

          5.8.2. Adjust the value of aperture diaphragm to 70%

80% of the value of objective’s N.A. 

          5.8.3 The aperture diaphragm could control the N.A. of condenser. Do not adjusting the brightness by 

it while use the brightness adjusting knob to make the adjustment 

          5.8.4 N.A. is marked in each objective 
                    e.g. 40X/0.65 stands for that magnification is 40X and N.A. is 0.65 
          5.8.5 Remove the eyepiece and observe through the eye tubes by eyes directly   
                    N.A.  of  condenser  could  tell  users  the  position  of  aperture  diaphragm  adjusting  shaft  in 

corresponding magnification. (That is to say, the value of aperture diaphragm should be 70% to 
80%  of  the  N.A.  of  objective  when  the  aperture  diaphragm  adjusting  shaft  reaches  some 
position) If the objective has been changed, please move the shaft to the position with the same 
value which is in the objective in the optic path and this will make ideal contrast 

 
 
 
 
 
 
 
  5.9 Observing with oil immersion 
          There is “oil” marked in the oil immersion objective. Please use the oil immersion supplied by the 

manufacturer between the objective and cover glass 

          5.9.1 Operation 
                      5.9.1.1 Condenser 
                                    Remove the slide back and fall the condenser down   
a little. One drop of oil could be placed in the top of condenser by through   
the long aperture of stage. Then move the slide forward and lift the condenser   
up 
                      5.9.1.2 Objectives 
                                    Turn the nosepiece and take the objective out of the   
optic path. Place one drop of oil on the slide then turn the condenser slowly   
and mount the objective 
        5.9.2 Air bubble in oil immersion 
                  5.9.2.1 Never make the oil immersion into your eyes. Please take following steps as soon as the 

oil immersion touches your skin or eyes 

                                Use soap or clean water to wash your skin carefully 
                                Please  wash  your  eyes  with  clean  water  (at  least  wash  15  minutes)  and  go  to  the 

hospital at once 

                  5.9.2.2 Do not expose the oil immersion in the sun or ultraviolet radiation directly 

Содержание A12.1010

Страница 1: ...A12 1010 Laboratory Biological Microscope Instruction Manual Please read this manual carefully before use microscope...

Страница 2: ...Contents BM2000 1 Structure and nomenclature 4 2 Technical parameter and specification 6 3 Configuration 8 4 Assemblage 9 5 Operation 13 6 Troubleshooting 18 7 Maintenance 20 8 Products Standards 20...

Страница 3: ...oscope on a stable plane and keep it in balance Keep the environment breezy and cover the microscope with the dust cover 2 6 Do not place microscope in moist room to avoid short circuit Please turn of...

Страница 4: ...inserted into the holder firmly Such Operation may cause short circuit or smoking At last please make sure if the collector has been mounted properly 9 Temperature for illumination There will be high...

Страница 5: ...re screw Nosepiece Objectives Specimen holder Mechanical stage Condenser with auxiliary lens Collector Binocular head Eye tubes fixture Microscope body Power supply Lightness adjusting knob switch scr...

Страница 6: ...Working environment Temperature 0 40 Humidity max 85 no dew Altitude within 2000m Pollution 2 Assembly 2 Optical parameter 2 1 Objectives infinity achromatic objectives Magnification N A Cover glass t...

Страница 7: ...tem Content Piece 1 Microscope s body including stage nosepiece coaxial coarse and fine focusing adjustment bracket for condenser adjusting power supply halogen bulb and normal collector 1pc 2 Siedent...

Страница 8: ...ded AC connector and make sure that the electronic cord has been connected safely Please use the supplied electronic cord by manufacturer Please make sure to use PE electronic cord when using extra lo...

Страница 9: ...rsion image disappeared 4 5 2 Objectives The objectives have been mounted in the microscope before shipment The objectives have been mounted in the microscope before shipment Take off the specimen fro...

Страница 10: ...lacement for fuse 4 6 2 1 Turn off the power supply turn the switch to O and pull out the connector 4 6 2 2 Open the cover for fuse by screwdriver 4 6 2 3 Mount the new fuse 4 6 2 4 Make sure that the...

Страница 11: ...there is sharp image 5 4 4 Turn the fine focusing knob to focus finely Use the 10X objective or 4X objective firstly before 40X and 100X objective Turn the fine focusing knob to focus finely 5 5 Adjus...

Страница 12: ...ctive when the aperture diaphragm adjusting shaft reaches some position If the objective has been changed please move the shaft to the position with the same value which is in the objective in the opt...

Страница 13: ...4 2 Never press the bottle overly as it may caused oil gushed out of the bottle 5 9 4 3 Clean the out surface of the bottle from remanent oil 5 10 Adjusting the coarse focusing knob s tension The tens...

Страница 14: ...or dust on the slide Cleaning Condenser position too low Correct position Poor image quality low resolution poor contrast No cover glass on the specimen Attach cover glass Cover glass too thick or thi...

Страница 15: ...Inadequate brightness or illumination Adjust brightness with control dial 3 Electrical Trouble Cause Remedy Lamp dose not light when switched ON No electrical power Check power cord connection Lamp b...

Страница 16: ...er thinner ect to clean the painted or plastic surfaces of the instrument We recommend the use of silicon cloth More persistent dirt may be cleaned with mild detergent solution Printed plastic surface...

Отзывы: